Development of solid reinforced porous femoral components made using electron beam powder bed fusion to meet endurance requirement
Yawen Huang, Philipp Nieke, Zhan Wen Chen
Key points
From the paper's conclusion- The present findings may suggest how highly porous EMPBF femoral components can be designed, built and quality assured to satisfy the fatigue life requirement.
Abstract
Solid stemmed femoral components are currently used, although they can be porous-coated. A highly porous component can enhance the long-term fixation of the implant through bone in-growth. Electron beam powder bed fusion (EBPBF) is capable of fabricating porous structures. However, a femoral component also needs to function under a highly critical loading condition. In this study, design/simulation, EBPBF and fatigue testing were conducted in order to develop satisfactory and highly porous Ti6Al4V femoral components. Loading during testing has closely followed that set by the international standard for femoral components. After the initial design, EBPBF and testing, a final design was made through topological optimisation and components were made. Tests have shown that two components have satisfied the requirement well (cycle number N > 5 x10 7 ) and two have failed each with a failure cycle number (N f ) slightly lower in logarithmic scale than the required (N f >5 x10 6 cycles). Then, an attempt has been made to explain the effect of EBPBF defect on N f using the Kitagawa-Takahashi (K-T) diagram approach through (a) tests to determine the threshold stress intensity factor, (b) fractography and micro-CT to locate the defects and measure the defect sizes, (c) close evaluation of stress in defect locations by simulation. It has been shown that the effect of defects affecting N f in a real component could approximately, although not highly accurately, be explained using the K-T diagram approach. The present findings may suggest how highly porous EMPBF femoral components can be designed, built and quality assured to satisfy the fatigue life requirement.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Importance of surface defect removal and compressive residual stress introduction in improving fatigue strength of powder bed-fused Ti6Al4V and demonstration of fatigue strength beyond that of hot-rolled Ti6Al4VInternational Journal of Fatigue · 2025
- Processing and Development of Porous Titanium for Biomedical Applications: A Comprehensive ReviewJournal of Manufacturing and Materials Processing · 2025
- Corrosion and corrosion fatigue studies of PBF-EB/M-manufactured porous iron in simulated body fluidnpj Materials Degradation · 2026
- Computed tomography based investigation of critical defects and fatigue failure analysis of very high cycle fatigued Ti-6Al-4V produced by laser powder bed fusionInternational Journal of Fatigue · 2026
- Computational Performance Evaluation of Porous Ti-6Al-4V Scaffold Architectures for Femoral Defect Healing Using a Coupled Finite Element-Mechanoregulation Framework.Journal of biomedical materials research. Part B, Applied biomaterials · 2026
- Finite Element Analysis of Conventional Fixation and 3D ‐Printed Scaffold Integration for Treating Large Osseous Femoral DefectsJournal of Biomedical Materials Research Part B Applied Biomaterials · 2025