Adam Schlauch, Bo He, Abhinav Nalluri, Ishan Shah, John Duldner, Tien Tran, Maria Caicedo, Oluwatodimu Richard Raji, Brian Farrell
This biomechanical study suggests that hip conditions are not significantly altered in T-type acetabular fractures outside the weight-bearing dome when there is an articular gap but no step-off and when there is an articular step-off and a gap of 2 mm or less. Infratectal T-type acetabular fractures with a combination of 2 mm articular step-off and gap greater than 5 mm may benefit from surgical reduction as malunion may lead to increased contact pressures within the joint.
PURPOSE: To determine the effect of gap and step-off displacement of infratectal T-type acetabular fractures in a biomechanical model and understand the effects of displacement on acetabular joint pressures when the fracture lies outside of the weight-bearing dome.
METHODS: Infratectal T-type acetabular fractures were created with osteotomies in cadaveric acetabuli and the hip joints were assembled in a servohydraulic test frame. Intra-articular pressure sensors were used to measure contact area, contact pressure, and peak forces through increasing gap and step-off displacement during simulated positions of standing, walking, and stair climbing.
RESULTS: With 0 mm articular step-off and increasing gap from 0 mm to 5 mm, there was no significant change in contact area, contact pressure, or peak force throughout most testing conditions. With a 2 mm step-off, contact area (-5.9 ± 1.8cm2, -23.9% ± 7.3%, p = 0.002) and contact pressures (10.9 ± 4.2 N/cm2, 43.5% ± 15.8%, p = 0.004) changed significantly once a 5 mm gap was reached.
CONCLUSION: This biomechanical study suggests that hip conditions are not significantly altered in T-type acetabular fractures outside the weight-bearing dome when there is an articular gap but no step-off and when there is an articular step-off and a gap of 2 mm or less. Infratectal T-type acetabular fractures with a combination of 2 mm articular step-off and gap greater than 5 mm may benefit from surgical reduction as malunion may lead to increased contact pressures within the joint.