Reproducibility of T2 relaxation time mapping in a phantom and human knee articular cartilage.

Henri P P Leskinen, Nea Pulkkinen, Antti Aarnio, Amir Esrafilian, Teija Sainio, Eveliina Lammentausta, Ville Kantola, Victor Casula, Simo Ojanen, Jari Rautiainen, Petri Kaurola, Mimmi K Liukkonen, Jarkko Iivarinen, Mikko J Nissi, Jari Heikkinen

一句话结论 · In one sentence

Low-to-moderate inter-scanner variation for the knee ROIs, CV = 2.2% to 22.8%, CVRMS,1.5 T = 9.3%, CVRMS,3 T = 11.0%, was observed. The phantom measurements showed low inter-scanner variation (CVRMS, 1.5 T = 4.7%, CVRMS, 3 T = 2.4%) on the typical cartilage T2 range.

原始摘要(英文原文)· Original abstract
OBJECTIVE: To quantify the reproducibility of knee articular cartilage T2 across a wide range of MR systems. MATERIALS AND METHODS: The reproducibility of T2 relaxation time mapping was assessed by measuring the knee of a volunteer (n1.5 T = 22, n3T = 13) and a System Standard 130 phantom (n1.5 T = 19, n3T = 12) in 38 different clinical MRI scanners at 16 imaging sites. Root-mean-square coefficient of variation (CVRMS) was used to quantify the reproducibility of T2 at the selected knee cartilage regions of interest (ROI) and the phantom T2 array vials. RESULTS: Low-to-moderate inter-scanner variation for the knee ROIs, CV = 2.2% to 22.8%, CVRMS,1.5 T = 9.3%, CVRMS,3 T = 11.0%, was observed. The phantom measurements showed low inter-scanner variation (CVRMS, 1.5 T = 4.7%, CVRMS, 3 T = 2.4%) on the typical cartilage T2 range. DISCUSSION: Reproducibility of quantitative T2 can vary between healthcare facilities and scanners but also between the cartilage compartments. The variation is sufficiently small to consider T2 mapping reproducible. Phantom measurements support this result.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Reproducibility of T2 relaxation time mapping in a phantom and human knee articular cartilage. — 科研速览 Science Skim