Temporal Dynamics and Clinical Relevance of Oscillatory Amplitude Variability During High-Frequency Oscillatory Ventilation With Volume Guarantee in Preterm Infants.
Kamal Ali, Mesaed Alsenani, Saad Alshareedah, Mohamed Sufyani, Saleh Algarni, Piero Alberti, Abdulaziz Homedi, Saif Alsaif,
Neonatal Intensive Care Department, King Abdulaziz Medical City, Riyadh, Saudi Arabia. · King Abdullah International Medical Research Centre, Riyadh, Saudi Arabia.
Key points
From the paper's conclusion- Single-centre retrospective cohort study using minute-level ventilator data collected during the first 72 h of HFOV-VG in 60 preterm infants.
- Amplitude CV and RMSSD declined over time. Variability was predominantly within-infant, while mean amplitude, VThf/kg, and frequency were predominantly between-infant. Within-infant increases in oscillatory frequency were associated with higher variability.
- Early oscillatory amplitude variability during HFOV-VG reflects dynamic ventilator-lung interactions and showed moderate discrimination for adverse respiratory outcome, requiring validation in larger independent cohorts.
Abstract
OBJECTIVE: To characterise temporal dynamics, variance structure, determinants, and predictive relevance of oscillatory amplitude variability during high-frequency oscillatory ventilation with volume guarantee (HFOV-VG) in preterm infants. METHODS: Single-centre retrospective cohort study using minute-level ventilator data collected during the first 72 h of HFOV-VG. Amplitude variability was quantified using standard deviation, coefficient of variation (CV), and root mean square of successive differences (RMSSD). Temporal patterns, variance decomposition, and determinants of variability were assessed using mixed-effects modelling separating within- and between-infant effects. Predictive performance for death before discharge or moderate-to-severe bronchopulmonary dysplasia was evaluated using receiver operating characteristic analysis. RESULTS: A total of 237,790 measurements from 60 infants were analysed. Amplitude CV declined from 0.122 (0.072-0.160) at 0-6 h to 0.063 (0.035-0.106) at 48-72 h, while RMSSD decreased from 2.10 (1.22-2.69) to 1.13 (0.64-1.94). Variability was predominantly within-infant (ICC 0.246-0.282), whereas mean amplitude, VThf/kg, and frequency were predominantly between-infant. Within-infant increases in oscillatory frequency were associated with higher variability, whereas increases in mean airway pressure and time from HFOV initiation were associated with lower variability. The composite outcome occurred in 47/60 infants. Early oscillatory amplitude variability demonstrated moderate discrimination (AUC 0.72, 95% CI 0.55-0.86), with an AUC of 0.82 (95% CI 0.66-0.95) when combined with oscillatory frequency and weight-normalised DCO2. CONCLUSIONS: Oscillatory amplitude variability during HFOV-VG reflects dynamic ventilator-lung interactions. Its association with adverse respiratory outcome is hypothesis-generating and requires validation in larger independent cohorts.
Read the paper
Discussion · 1
Science Skim's AI key points, plus readers' public thoughts
- Single-centre retrospective cohort study using minute-level ventilator data collected during the first 72 h of HFOV-VG in 60 preterm infants.
- Amplitude CV and RMSSD declined over time. Variability was predominantly within-infant, while mean amplitude, VThf/kg, and frequency were predominantly between-infant. Within-infant increases in oscillatory frequency were associated with higher variability.
- Early oscillatory amplitude variability during HFOV-VG reflects dynamic ventilator-lung interactions and showed moderate discrimination for adverse respiratory outcome, requiring validation in larger independent cohorts.
Readers' thoughts
No reader thoughts yet.
Related papers
- Physiological and clinical comparison of two high-frequency oscillatory ventilation with volume guarantee strategies in preterm infants with respiratory distress syndrome.Neonatology · 2026
- Cardiorespiratory coupling during high-frequency oscillatory ventilation with volume guarantee in extremely preterm infants.European journal of pediatrics · 2026
- 31 An examination of high frequency oscillatory ventilation and its impact on respiratory outcomes in extremely low gestational age infants: A retrospective analysisPaediatrics & Child Health · 2026
- Volume Guarantee Versus Manual Tidal Volume Monitoring During High‐Frequency Oscillatory Ventilation in Neonates: A Pilot Randomized Controlled TrialPediatric Anesthesia · 2026
- High-Frequency Oscillation vs Mechanical Ventilation for Neonatal Acute Respiratory Distress SyndromeJAMA Network Open · 2026
- Explained and Residual Variability of the Fetal Mechanical PR Interval in Structurally Normal Hearts: A Secondary Analysis of Physiological, Biometric, and Doppler Covariates.Biomedicines · 2026