Histological Investigations and Comparative Analysis of a Novel 455 nm Blue Light Laser.
Maxwell Scher, Margaret Mills, Denis Lafreniere, Gregory A Grillone, Lauren F Tracy
Department of Otolaryngology-Head and Neck Surgery, Boston University Chobanian & Avedisian School of Medicine, Boston, Massachusetts, USA. · Division of Otolaryngology - Head and Neck Surgery, University of Connecticut School of Medicine, Farmington, Connecticut, USA.
Key points
From the paper's conclusion- The 455 nm laser demonstrated similar incision depths to the 10,600 nm CO2 laser at 5 and 10 W, with increasing depth at 15 and 20 W. The 455 nm laser averaged 0.32 mm greater thermal damage depth than the CO2 laser. Increasing power for 455 nm laser resulted in deeper incisions without a proportional increase in thermal damage, indicating potential use as a cutting laser.
Abstract
OBJECTIVES: To histologically evaluate the cutting effects of a novel 455 nm laser across a range of power and distance settings and to compare with a 10,600 nm fiber-based carbon dioxide (CO2) laser. STUDY DESIGN: Descriptive study. SETTING: In vitro. METHODS: Using the 455 nm blue laser, and 10,600 nm CO2 laser, a 300 µm fiber was secured at specific distances (range: 1-5 mm) from ex vivo bovine tongue tissue. The tissue was incised at a constant speed and varying power (5-20 W). Histologic specimens were stained for hematoxylin and eosin (H&E) and lactate dehydrogenase (LDH). Incision depth and thermal damage were measured and compared across different wavelengths, power, and distances from tissue. RESULTS: At 5 W, the average incision depths were 0.448 mm for 455 nm laser and 0.643 mm for 10,600 nm laser (P = .087). At 10 W, average incision depth was 1.01 mm for 455 nm laser and 1.31 mm for 10,6000 nm laser (P = .090). Increasing power to 20 W increased the average incision depth for the 455 nm laser to 1.588 mm (P < .0001). Laser distance from tissue (1-5 mm) did not significantly impact incision depth. Thermal damage depth of the 455 nm laser averaged 0.470 mm versus 0.150 mm for the 10,600 nm laser (P < .0001). Power adjustments did not significantly alter thermal spread depth for either laser. CONCLUSION: The 455 nm laser demonstrated similar incision depths to the 10,600 nm CO2 laser at 5 and 10 W, with increasing depth at 15 and 20 W. The 455 nm laser averaged 0.32 mm greater thermal damage depth than the CO2 laser. Increasing power for 455 nm laser resulted in deeper incisions without a proportional increase in thermal damage, indicating potential use as a cutting laser.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Ex vivo histological comparison between Er, Cr: YSGG laser and slow-speed hand piece in bone cutting.Wiadomosci lekarskie (Warsaw, Poland : 1960) · 2026
- Temporal Histological Tissue Responses Following Fractional Nonablative 1470 nm Laser Exposure: A Pilot In Vivo Human Study.Aesthetic surgery journal. Open forum · 2026
- Thermal Tissue Damage and Healing After Laser Excision of Benign Oral Soft-Tissue Lesions: An Optical Coherence Tomography (OCT)-Based Comparison of 980-nm Diode and 1064-nm Nanosecond-Pulsed Nd:YAG Laser.Lasers in surgery and medicine · 2026
- Medium Power Density Blue Laser Photocoagulation for Vocal Fold Vascular Lesions.Journal of voice : official journal of the Voice Foundation · 2026
- Ablation Depth and Surface Morphology of Dentin After Irradiation with a Tunable TEA-CO2 Laser in the Wavelength Range of 9.3 - 10.6 µm.Journal of lasers in medical sciences · 2026
- Effect of laser pulse width (ns-fs) on laser-induced damage thresholds of a CMOS photodetector.Applied optics · 2026