Low‐Cost, Computer‐Free Spin Echo Generator Board
Helina K. Wilson, Christopher E. Vaughn, Antonio Glenn, Jessie E. P. Sun, Jason P. Stockmann, Lawrence Wald, William A. Grissom, Sai Abitha Srinivas
Key points
From the paper's conclusion- Inspired by earlier work on inexpensive spin echo generator circuit boards, the board uses low‐cost digital timing circuits for pulse generation and mixers for RF pulse gating, without requiring a computer.
Abstract
A Spin Echo Generator Board has been designed as an affordable tool for MR system quality assurance, demonstration, and education. Inspired by earlier work on inexpensive spin echo generator circuit boards, the board uses low‐cost digital timing circuits for pulse generation and mixers for RF pulse gating, without requiring a computer. It produces a simple rectangular excitation pulse followed by a refocusing pulse, with user‐adjustable pulse widths, TE, and TR delays. The board features an integrated T/R switch, pre‐amplifier, and an onboard 9‐V battery holder for standalone operation. External requirements include only a synthesizer, RF power amplifier, oscilloscope for display, and a tuned RF coil placed within a magnet. The hands‐on platform and real‐time feedback support practical applications, such as verifying signal generation and coil performance. In an educational setting, the board enables students to explore how variations in timing parameters influence MR signal behavior. It provides a transparent learning environment in which students can investigate fundamental MRI concepts, including flip angle, resonance, and T 2 relaxation, through direct experimentation. Functionality is demonstrated on a 0.5‐T tabletop magnet and 0.1‐T whole‐body MRI scanner.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Concurrent excitation and detection in magnetic resonance using voltage-controlled oscillators.Science advances · 2026
- A Low-Cost MR-Compatible Pneumatic Motion Phantom for Magnetic Resonance Imaging.ArXiv · 2026
- Embedded Trigger Control System for Wave--Plasma Interaction Experiments in Large Volume Plasma DevicearXiv · 2026
- SeqEyes: An Open-Source Toolbox for Synchronized Waveform/Trajectory Visualization in Pulseq.Magnetic resonance in medicine · 2026
- An ECG-PPG Physiological Signal Emulator for Calibration and Validation of Cardiovascular Monitoring Devices.Sensors (Basel, Switzerland) · 2026
- A Software-Centered Virtual–Physical Integrated Experimental Platform for Synchronous Generator EducationElectronics · 2026