B 1 + Optimization in the Heart at 7 T.
Erin McConnell Montoya, Aaron T Hess, Damian J Tyler, Ladislav Valkovič
Oxford Centre for Clinical MR Research (OCMR), RDM Cardiovascular Medicine, University of Oxford, Oxford, UK. · Oxford Centre for Integrative Neuroimaging (OxCIN), Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, UK.
Key points
From the paper's conclusion- A fixed, population-derived, global phase shim provides a practical, calibration-free approach for improving cardiac B 1 + performance at 7 T. The present results extend previous universal RF-shimming work through prospective evaluation in an independent validation cohort and demonstrate that much of the benefit of personal shimming can be retained without participant-specific optimization. The resulting fixed-phase configuration could also provide a practical route towards implementation on single-transmit hardware for cardiac imaging and spectroscopic applications.
Abstract
PURPOSE: At 7 T, the short RF wavelength in tissue (∼11 cm) causes destructive interference within the chest, leading to B 1 + inhomogeneity and posing a critical challenge for cardiac MRI and MRS. While subject-specific parallel-transmit (pTx) optimization can mitigate this, it requires calibration scans and offline computation. To simplify this process, a static "global" shim was calculated and demonstrated to outperform the default transmit configuration without participant-specific calibration. METHODS: B 1 + maps were acquired in 12 healthy volunteers at 7 T using an 8-channel 1H torso coil. "Personal" phase-only shims were calculated per subject by minimizing a cost function of B 1 + homogeneity and efficiency. The "global" shim was calculated by optimizing performance across the cohort, then validated in six additional volunteers by comparing performance against the native mode. RESULTS: In the validation cohort, the "global" significantly outperformed the native shim. It increased the Mean Transmit Sensitivity by ∼20% and reduced CoV by ∼ 16 % in the septum. The fixed global shim performance was within 3% of the computationally expensive personal shim for all metrics. CONCLUSION: A fixed, population-derived, global phase shim provides a practical, calibration-free approach for improving cardiac B 1 + performance at 7 T. The present results extend previous universal RF-shimming work through prospective evaluation in an independent validation cohort and demonstrate that much of the benefit of personal shimming can be retained without participant-specific optimization. The resulting fixed-phase configuration could also provide a practical route towards implementation on single-transmit hardware for cardiac imaging and spectroscopic applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of Amplitude- and Amplitude-Frequency-Modulated Pulses for Outer Volume Suppression in Cardiac MRI: Application to Myocardial T 2 Mapping.Magnetic resonance in medicine · 2026
- High Dynamic Range Turbo-FLASH B1 Mapping for Characterizing Multi-Transmit Multi-Receive RF Arrays, Validated at 7 T and 11.7 T.Magnetic resonance in medicine · 2026
- Design of an 8‐Channel Transmit 32‐Channel Receive 11.7T Head Coil and Evaluation of SNR GainsMagnetic Resonance in Medicine · 2026
- Evaluation of 7 Tesla pTx Body Coils Regarding Transmit and Receive Performance.Magnetic resonance in medicine · 2026
- Integrating Low-loss High-permittivity Ceramic Helmet with RF Head Array Coil to Largely Enhance Coil Sensitivity and SNR for Human Brain MRI at 7T.IEEE transactions on bio-medical engineering · 2026
- Whole-Brain Diffusion MRI at 7T With Universal Multiband GRAPE pTx Pulses.Magnetic resonance in medicine · 2026