Independent Component Analysis for Non-Contrast-Enhanced Free-Breathing Pulmonary Ventilation and Perfusion MRI.
Laura Rozo Pardo, Rabea Klaar, Enrico Schulz, Moritz Rabe, Christopher Kurz, Guillaume Landry, Jens Ricke, Bastian Sabel,
Department of Radiology, LMU University Hospital, LMU Munich, Munich, Germany. · Department of Radiation Oncology, LMU University Hospital, LMU Munich, Munich, Germany.
Key points
From the paper's conclusion- ICA functional lung MRI provides a robust alternative to conventional techniques, enabling reliable ventilation and perfusion assessment from free-breathing acquisitions, particularly in the presence of irregular respiratory patterns and motion.
Abstract
PURPOSE: To introduce a robust approach for non-contrast-enhanced free-breathing functional lung MRI based on independent component analysis (ICA), addressing key limitations of standard methods such as sensitivity to irregular breathing or irregular heartbeat. METHODS: Repeated dynamic single-slice MRI of ten healthy volunteers and twenty-five lung tumor patients was performed in free breathing on a 0 . 35 T MR-linac using a 2D coronal cine bSSFP sequence, resulting in 143 datasets. Following motion correction, lung segmentation, and signal alignment, temporal ICA was applied to extract ventilation and perfusion components. Ventilation- and perfusion-weighted maps were generated and compared to Fourier decomposition (FD) and matrix pencil (MP) analysis. Signal construction performance was evaluated using the Pearson correlation coefficient and nonparametric statistical testing, comparing the methods under regular and irregular breathing and heartbeat conditions. Map accuracy and robustness at low signal-to-noise ratio (SNR) were additionally assessed in numerical-phantom simulations. RESULTS: In vivo, ICA effectively separated physiological lung signals, achieving significantly higher temporal agreement with original voxel time courses than both FD and MP (all p < 0 . 001 ). Under all physiological conditions, ICA yielded the highest temporal correlations. In simulations, ICA maintained robust signal fidelity, especially under irregular breathing and low-SNR conditions, preserving > 90 % of contrast in simulations, while FD and MP exhibited pronounced degradation and significantly reduced temporal correlations. CONCLUSION: ICA functional lung MRI provides a robust alternative to conventional techniques, enabling reliable ventilation and perfusion assessment from free-breathing acquisitions, particularly in the presence of irregular respiratory patterns and motion.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Adapted Linear Binning Method to Assess Pulmonary Ventilation and Perfusion in Children Using Lung Matrix Pencil Decomposition (MP)-MRI.Magnetic resonance in medicine · 2026
- Evaluating Perfusion Quantification for Phase‐Resolved Functional Lung MRI Using Gaussian FilteringNMR in Biomedicine · 2026
- Respiratory rate estimation by using realignment data from a 1-minute high-temporal-resolution MRI sequence: A feasibility study.Acta radiologica (Stockholm, Sweden : 1987) · 2026
- Early post-transplant ventilation and perfusion assessment using phase-resolved functional lung MRI.La Radiologia medica · 2026
- Estimating pediatric lung volumetric parameters via rapid, limited-slice, free-breathing thoracic dynamic MRI.Medical physics · 2026
- Dynamic DWI detects respiratory modulation of brain parenchymal hydrodynamics.Imaging neuroscience (Cambridge, Mass.) · 2026