Bias-corrected respiratory motion standard deviation from amplitude-binned 4D-MRI for radiotherapy planning.
Michael Ferraro, Sirisha Tadimalla, Fidel Navarro Salazar, Eireann Cosgriff, Annette Haworth, Jonathan Sykes
Institute of Medical Physics, University of Sydney, Sydney, NSW, Australia. · Radiation Oncology Network, Western Sydney Local Health District, Sydney, NSW, Australia.
Key points
From the paper's conclusion- Conclusions The proposed estimator provides an accurate measurement of respiratory motion SD from amplitude‐binned 4D‐MRI by accounting for intra‐bin variability.
Abstract
BACKGROUND: Respiratory-resolved magnetic resonance imaging (4D-MRI) provides superior soft-tissue contrast for characterizing respiratory motion in radiotherapy, but its geometric fidelity depends on the reconstruction and binning strategy. As with 4D-CT, peak-to-peak amplitude derived from binned 4D-MRI is prone to underestimation. However, as acquisitions span many respiratory cycles, the standard deviation (SD) of motion, required for probability-based planning approaches such as mid-ventilation, may be estimated accurately, provided binning-induced bias is corrected. PURPOSE: To develop and validate a motion SD estimator for equal-probability amplitude-binned 4D-MRI data that corrects for systematic underestimation caused by intra-bin motion variability. METHODS: 4D-MRI data were acquired on a dynamic motion phantom driven by five patient-derived respiratory traces. Images were reconstructed using both a standard non-uniform fast Fourier transform (NUFFT) and the compressed sensing-based XD-GRASP algorithm with varying numbers (2-20) of equal-probability respiratory amplitude bins. An SD estimator based on the law of total variance was implemented to correct for intra-bin variance. The accuracy of peak-to-peak amplitude and SD was evaluated against ground-truth physical phantom motion, with performance under deformable motion tested on a digital phantom (CoMBAT). RESULTS: Both reconstruction methods underestimated true peak-to-peak motion. While the inter-bin SD calculation also underestimated motion variability, the proposed correction substantially mitigated this bias. It yielded robust SD estimates with median residual underestimation of 0.13 mm (5 bins) and 0.11 mm (10 bins) across patient traces. In CoMBAT, the absolute bias of the total SD estimate was below 3% from five bins onward. CONCLUSIONS: The proposed estimator provides an accurate measurement of respiratory motion SD from amplitude-binned 4D-MRI by accounting for intra-bin variability. This supports the reliability of 4D-MRI for planning strategies dependent on motion probability distributions. Additionally, the confirmed underestimation of peak-to-peak amplitude highlights a fundamental limitation in defining internal target volumes directly from binned 4D-MRI.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Protocol optimization for quantitative MRI studies in radiation oncology: II. Diffusion MRIPhysics in Medicine and Biology · 2026
- DoReMe: Dual optimization of Reconstruction and Motion estimation for self-navigated free-breathing abdominal MRI using Gaussian SplattingarXiv · 2026
- Sensitivity of Pulmonary 4D Flow MRI Biomarkers to Controlled Spatial and Temporal Degradation in a CFD BenchmarkbioRxiv · 2026
- Protocol optimization for MRI studies in radiation oncology: III. MRI relaxometryPhysics in Medicine and Biology · 2026
- Prior-Guided Physics-Informed Deep Learning for Fast 4D MRI ReconstructionarXiv · 2026
- Effect of Visualized Respiratory Training on Patient Cooperation and Image Quality in Gadoxetic Acid Disodium‐Enhanced Liver MRI : A Randomized Controlled TrialJournal of Magnetic Resonance Imaging · 2026