Assessing effects of low-dose CT attenuation correction on quantification and image quality on a LAFOV PET system with a modified NEMA-IQ phantom.
Vanessa Nadig, Jane Mackewn, Georgios Krokos, Lewis C Davies, Daniel Darian, Sarah-May Gould, Gary J R Cook, Alexander Hammers,
Siemens Healthcare Limited, Camberley, UK. · King's College London and Guy's & St Thomas' PET Centre, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
Key points
From the paper's conclusion- Even for low-dose CT protocols, PET quantification remained consistent in all NEMA-IQ spheres for many materials. Beam hardening and photon starvation in the CT image are likely to occur in specific body regions such as hips, shoulders or the abdomen for an arms-down scan and these effects had a bigger impact on the quantification than any difference in HU in both soft tissue and bone. This therefore should be considered when selecting suitable CT imaging parameters.
Abstract
BACKGROUND: Over the past decades, efforts have been directed toward increasing the sensitivity of PET systems to enable low-dose PET imaging and shorter scan times. The Biograph Vision Quadra PET/CT scanner (Siemens Healthineers, Knoxville, TN, USA) combines an increased solid angle coverage with time of flight, enabling a significant reduction in administered activity. To fully benefit from low-dose PET imaging, the dose of the CT scan should ideally also be reduced. PURPOSE: The CT radiation dose can be reduced by lowering the tube voltage and exposure as well as by the use of tin filtration. Reducing the CT dose must be carefully aligned with any impact that changing CT image quality and quantitative metrics may have on the PET quantification, for example via linear attenuation coefficients. We evaluate the impact of low-dose CT protocols on the PET quantification in soft-tissue-like and bone-equivalent materials using a customized phantom. METHODS: The phantom was assembled out of a NEMA-IQ phantom, several layers of saline and two bovine long bones, to simulate a patient positioned with arms down. A 10-min PET scan with 47 different accompanying CT scans was acquired. The CT tube voltage and exposure were varied between 70 k V to 140 k V , with and without tin filtration, and 7 mAs to 400 mAs , respectively. The effect of the different CT parameters on the PET image quality and quantitative accuracy was assessed via several quantitative metrics, including standardized uptake value (SUV) and signal-to-noise ratio, along with the CT dose length product (DLP). RESULTS: The recorded DLP values suggest that using all tube voltages except 140 k V combined with exposures in the range of 7 m A s to 30 m A s , can result in dose savings of ≥ 80 % compared to a typical clinical protocol ( 120 k V , 50 mAs ). Compared to this typical protocol, HU differed more in bone ( 10 % to 40 % difference) than in soft tissue ( 1 % to 30 % difference), and more streak artefacts and photon starvation were seen when configuring low tube voltages and exposures, with an accompanying increase in CT noise. The observed artefacts were transferred with low intensity to the PET images via the attenuation map. The SUV mean remained within 2 % of the SUV mean of the typical clinical protocol for most tested protocols. The presence of artefacts in the CT images had a bigger impact on PET quantification in both soft tissue and bone ( SUV mean difference of 10 % to 20 % ) than any difference in HU. Protocols that resulted in artefacts showed an increased inaccuracy in attenuation correction and scatter estimation, as assessed via the "lung" residual error (up to 10 % to 20 % increase). CONCLUSIONS: Even for low-dose CT protocols, PET quantification remained consistent in all NEMA-IQ spheres for many materials. Beam hardening and photon starvation in the CT image are likely to occur in specific body regions such as hips, shoulders or the abdomen for an arms-down scan and these effects had a bigger impact on the quantification than any difference in HU in both soft tissue and bone. This therefore should be considered when selecting suitable CT imaging parameters.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Image quality and radiation dose optimization in CT pulmonary angiography using 120, 100, and 80 kV protocols for pulmonary thromboembolism diagnosis.Quantitative imaging in medicine and surgery · 2026
- Size-specific weighted detectability index for computed tomography image characterization.Medical physics · 2026
- Optimization of patient dose and image quality for positron emission tomography-computed tomography.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Quantitative comparison of 68-keV monoenergetic, routine-dose, and low-dose 120-kVp CT for calcium-mimicking inserts in thoracic phantoms: effects of phantom size and reconstruction method.European radiology experimental · 2026
- Impact of CT parameter reduction on quantitative SPECT/CT assessment in 99mTc imaging: A phantom study.Radiography (London, England : 1995) · 2026
- CT protocol optimisation for oncologic surveillance imaging: phantom-based comparison of multi-scan and single-scan acquisition strategies.Biomedical physics & engineering express · 2026