Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine

In Vitro and in Vivo Characterization of a Miniature Histotripsy Platform Developed for Endoscopic or Laparoscopic Cancer Applications.

Connor S Centner, Richard Zhang, Rahim Shalash, John T Moore, Omar Sadi Sarkar, Emily Schulman-Geltzer, Theresa Weis, Matthew Mallay,

University of Louisville School of Medicine, Louisville, Kentucky, USA. · Department of Biology, University of Louisville, Louisville, Kentucky, USA.

Preclinical

Key points

From the paper's conclusion
  • These findings support the feasibility of a miniature histotripsy device for tumor cell lysis and pro-inflammatory macrophage polarization in vitro and provide initial in vivo evidence of localized tumor disruption, supporting further investigation for endoscopic and laparoscopic applications.

Abstract

OBJECTIVES: Extracorporeal histotripsy is limited by poor acoustic transmission across air-tissue interfaces, restricting broader therapeutic applications. This study evaluates in vitro and in vivo feasibility of a miniature histotripsy device designed for endoscopic and laparoscopic oncology applications. METHODS: A custom-designed 10 mm histotripsy transducer was tested in vitro on MC38 and Neuro2a cells at 160-240 V, with the samples assessed after 1 and 2 minutes of cumulative treatment and compared with untreated controls. Subsequent studies in MC38, Neuro2a, A549, and MDA-MB-231 cells were performed at 240 V for 2 minutes. Cell viability was quantified using trypan blue exclusion assay and flow cytometry with propidium iodide staining. Histotripsy-derived MC38 lysate was used to assess RAW 264.7 macrophage polarization. Preliminary in vivo feasibility was assessed in bilateral MC38 murine tumors at 240 V for up to 90 minutes, with post hoc mean grayscale intensity analysis performed. RESULTS: In vitro studies demonstrated cytotoxic effects in MC38 and Neuro2a cells that varied with driving voltage and treatment duration. Macrophages exposed to histotripsy-derived MC38 lysate exhibited increased M1 polarization (p < .01). Additionally, MDA-MB-231 cells exhibited lower propidium iodide positivity than MC38 and Neuro2a cells (p < .01). Histotripsy-treated regions of MC38 murine tumors demonstrated lower mean grayscale intensity than untreated regions (p < .001). CONCLUSION: These findings support the feasibility of a miniature histotripsy device for tumor cell lysis and pro-inflammatory macrophage polarization in vitro and provide initial in vivo evidence of localized tumor disruption, supporting further investigation for endoscopic and laparoscopic applications.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers