Development of a Gas Chromatographic Method for Determination of CO2/CH4 Breakthrough Curves in Fixed-Bed Adsorption.
Tamille A Souza, Mirtes F S Freitas, Argenis A S Jimenez, Yanier S Hechavarria, Raildo A Fiuza-Júnior, Artur J S Mascarenhas, Karen V Pontes
Laboratório de Processos Sustentáveis e Energias Renováveis (LAPSER), Departamento de Engenharia Química, Universidade Federal da Bahia, Salvador, Brazil. · Grupo de Pesquisa em Bioanalítica e Metabolômica Clínica (BioMet), Departamento de Química Analítica, Instituto de Química, Universidade Federal da Bahia, Salvador, Brazil.
Key points
From the paper's conclusion- Overall, the proposed approach provides a robust, accessible, and easily implementable strategy for breakthrough curve determination using widely available GC configurations, offering a practical alternative for laboratories lacking fast-response analytical systems.
Abstract
Adsorption stands out as a promising alternative for CH4/CO2 separation of biogas and natural gas. The design of fixed-bed columns depends on laboratory-scale breakthrough curve assays, which require fast-response analytical methods such as microchromatographs or gas analyzers. However, these instruments are not always available in research laboratories, where conventional gas chromatography (GC) represents a more accessible alternative. Nevertheless, conventional GC generally does not achieve analysis times on the order of seconds. In this context, this work aims to develop and refine a chromatographic method based on conventional GC to obtain breakthrough curves for CO2/CH4 mixtures in a fixed-bed reactor, with reduced data-acquisition time. A chromatographic method was developed for constructing breakthrough curves using a two-level factorial design, achieving a chromatographic cycle time of approximately 4.5 min while maintaining adequate CO2/CH4 separation. The method demonstrated linearity within the evaluated concentration ranges, with low limits of quantification and detection, and was precise and robust. A sequential injection strategy was implemented to acquire chromatographic data during continuous-flow breakthrough experiments. The method was tested and evaluated with different adsorbents (zeolite and activated carbon), successfully reproducing CO2/CH4 breakthrough profiles and enabling the determination of dynamic adsorption capacity and selectivity. These results validate the proposed method's analytical performance and applicability under realistic separation conditions. Overall, the proposed approach provides a robust, accessible, and easily implementable strategy for breakthrough curve determination using widely available GC configurations, offering a practical alternative for laboratories lacking fast-response analytical systems.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance Evaluation of the Adsorption of a Binary Mixture (CH4/CO2) on Commercial Activated Carbons for Biogas UpgradingC · 2026
- Impact of CO₂ flushing and headspace volume on gas and methane kinetics in two in vitro Rumen systems: A comparative study of AnkomRF and gas endeavourAnimal Feed Science and Technology · 2026
- Dynamic activation catalysts for CO2 hydrogenationNature Communications · 2025
- Four-column Vacuum Pressure Swing Adsorption for biogas upgrading with co-production of high-purity biogenic CO2: Experimental and simulation studyJournal of CO2 Utilization · 2026
- Experimental Investigation of CO 2 /CH 4 Competitive Adsorption Behaviors in Shale under Different Feed Gas Injection MethodsEnergy & Fuels · 2026
- Experimental Investigation of CO2-CH4 Displacement Behavior during the Soaking Stage of CO2 Huff-and-Puff in Coal.ACS omega · 2026