JPEN. Journal of parenteral and enteral nutrition

Resting energy expenditure measured by whole-room and portable indirect calorimetry in adults with breast cancer: A comparison study.

Ana Paula Pagano, Taiara Scopel Poltronieri, Maria Cristina Gonzalez, William J Evans, Anil A Joy, Karen King, Claude Pichard, Carla M Prado

Department of Agricultural, Food, & Nutritional Science, University of Alberta, Edmonton, Alberta, Canada. · Postgraduate Program in Nutrition and Food, Federal University of Pelotas, Pelotas, Rio Grande do Sul, Brazil.

Observational

Key points

From the paper's conclusion
  • The Q-NRG demonstrated good group-level agreement and reliability for resting energy expenditure measurement. However, the wide variation suggests caution when interpreting individual results.

Abstract

BACKGROUND: Accurate measurement of resting energy expenditure is essential for informing nutrition support and optimizing outcomes. However, gold standard methods for measuring resting energy expenditure are often unfeasible in clinical settings, and previous portable devices have lacked accuracy. We investigated the accuracy and reliability of a portable device to measure resting energy expenditure in patients with breast cancer. We also compared its performance with that of commonly used predictive equations. METHODS: In this cross-sectional study, Q-NRG (COSMED, Rome, Italy) accuracy was assessed by calculating the proportion of resting energy expenditure measurements within 10% of those obtained by a whole-room indirect calorimeter and by using Bland-Altman analysis. Reliability was evaluated using the intraclass correlation coefficient, analyzing complete 10-min and last 5-min tests. RESULTS: Fifty females with breast cancer (mean ± SD, age: 56 ± 12 y; body mass index: 28.5 ± 7.2 kg/m2, stages II-III: 22%) were included. The Q-NRG accurately measured resting energy expenditure in ~70% of participants (~10% overestimations, ~20% underestimations). Individual-level agreement was better using complete test data (bias: -59 ± 124 kcal; limits of agreement: -301 to 183 kcal; absolute limits of agreement: 484 kcal). Good reliability was observed (intraclass correlation coefficient, complete tests = 0.895; last 5-min = 0.832). CONCLUSION: The Q-NRG demonstrated good group-level agreement and reliability for resting energy expenditure measurement. However, the wide variation suggests caution when interpreting individual results.

Read the paper