Journal of arrhythmia

Successful Bailout Retrieval of an Active-Fixation Atrial Leadless Pacemaker Following Suspected Tissue Entrapment During Repositioning.

Toshiya Yoshida, Akira Kasagawa, Daisuke Togashi, Ikutaro Nakajima, Tomoo Harada

Department of Cardiology St. Marianna University School of Medicine Kawasaki Japan.

Key points

From the paper's conclusion
  • Here, we report that advancing the protective sleeve to fully encapsulate the device potentially enabled safe and successful retrieval.

Abstract

An 81-year-old woman with a medical history of mechanical mitral valve replacement, Maze procedure, and multiple myeloma presented with symptomatic sinus arrest and underwent AVEIR DR implantation. Although the AVEIR VR was successfully implanted, optimal electrical parameters could not be achieved during AVEIR AR placement. During repositioning of the AVEIR AR, myocardial entrapment with helix stretching was encountered. Here, we report that advancing the protective sleeve to fully encapsulate the device potentially enabled safe and successful retrieval.