Integrated PRM and DIA via HILIC-Orbitrap HRMS for Simultaneous Quantification and Interference Screening of Amino Acids in Human Saliva.
Min Ju, Yuming Song, Yuming Sun, Lina Zhou, Jinfeng Zhao, Qingxin Yin, Huihui Wan
Instrumental Analysis Center, Dalian University of Technology, Dalian, China. · School of Chemical Engineering, Dalian University of Technology, Dalian, China.
Key points
From the paper's conclusion- In this study, a robust method based on hydrophilic interaction liquid chromatography coupled with Orbitrap high-resolution mass spectrometry (HILIC-Orbitrap HRMS) was developed for the simultaneous quantification of 20 amino acids in human saliva.
Abstract
Free amino acids in human saliva are increasingly recognized as valuable biomarkers for noninvasive health monitoring and disease diagnosis owing to their close association with systemic metabolic status. However, their accurate quantification remains challenging because of low physiological concentrations and high matrix complexity. In this study, a robust method based on hydrophilic interaction liquid chromatography coupled with Orbitrap high-resolution mass spectrometry (HILIC-Orbitrap HRMS) was developed for the simultaneous quantification of 20 amino acids in human saliva. Chromatographic separation was achieved using a BEH Amide column (150 mm × 2.1 mm, 1.7 μm) with a gradient elution program. Mass spectrometric detection was performed in positive-ion mode using a Full MS plus scheduled parallel reaction monitoring (PRM) acquisition scheme (hereafter referred to as PRM). The method exhibited excellent linearity (R2 > 0.991), accuracy (80.01%-119.60%), and precision (RSD < 10%). The limits of detection and quantification were 0.030-0.100 and 0.125-0.500 μmol/L, respectively. In addition, data-independent acquisition (DIA) was used to screen for co-eluting isobaric interferences; potential interferents such as 5-aminovaleric acid (for valine) and N,N-dimethylglycine (for 4-aminobutyric acid) were flagged and subsequently resolved with authentic standards. This validated HILIC-HRMS method provides a reliable platform for targeted quantification of polar amino acids in saliva, with DIA serving as a complementary tool for interference screening to ensure analytical specificity.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of a Hydrophilic Interaction Liquid Chromatography Design Space for Amino Acids.Journal of separation science · 2026
- Development and systematic evaluation of an untargeted HILIC‑HRMS metabolomics workflow for multi-matrix human samples analyses.Metabolomics : Official journal of the Metabolomic Society · 2026
- Amino acid profiles in exhaled breath condensate: evaluation of collection methods and diurnal variation Using LC-MS/MS.Journal of breath research · 2026
- Analysis of Histamine in Amino Acid Additives by Ultra-Performance Liquid Chromatography-Tandem Mass SpectrometryDOAJ (DOAJ: Directory of Open Access Journals) · 2026
- Simultaneous hydrophilic interaction liquid chromatography-tandem mass spectrometry quantification of serum neurotransmitters and related metabolites: Development, validation, and application to antenatal depression.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026
- Development and Validation of an LC–MS/MS Method for Simultaneous Determination of Gamma-Aminobutyric Acid (GABA), Glutamic Acid, and Glutamine in Fermentation SupernatantACS Food Science & Technology · 2026