A Survey and Analysis of the Current Status of MRI Safety Management in Medical Institutions in Korea.
Chung-Hwan Kang, Kyoo-Jin Yeon, Yu-Jin Shin, Kyung-Bae Lee
Department of Radiology, Konkuk University Medical Center, Seoul, Republic of Korea. · Department of Radiology, Samsung Medical Center, Seoul, Republic of Korea.
Key points
From the paper's conclusion- The response rate was 100% (47/47). The mean number of MRI units per hospital was 5.32 ± 3.50, and 3.0 T systems were predominant. Compliance was universal for closed-circuit television (CCTV) monitoring and MR Conditional monitoring devices but was low for ferromagnetic detection systems (17.0%), Magnetic Resonance Safety Officer (MRSO) appointment (17.0%), and Level 1 safety education (23.4%). MRI-related safety incidents were reported by 44.7% of hospitals. Two-person staffing decreased significantly from 97.9% on weekday daytime hours to 55.3% on Saturdays, 44.7% at night, and 38.3% on holidays. The overall composite safety maturity score was 63.08 ± 15.04.
Abstract
BACKGROUND: MRI safety management is essential for patient safety and for standardizing institutional practices. However, nationwide data on MRI safety readiness in Korean tertiary hospitals remain limited. PURPOSE: To evaluate MRI safety management across all 47 Korean tertiary hospitals and identify major gaps in safety infrastructure, staffing, and education. STUDY TYPE: Cross-sectional census survey. POPULATION: MRI safety managers from all 47 Korean tertiary hospitals in 2025. ASSESSMENT: A structured 50-item questionnaire was developed based on the American College of Radiology (ACR) MRI safety guidelines. A composite safety maturity index (0-100) was calculated as the unweighted arithmetic mean of four domain scores derived from 32 binary safety-practice items, with mandatory items weighted 2 and recommended items weighted 1 in accordance with the 2024 ACR Manual on MR Safety. STATISTICAL TESTS: Descriptive statistics were used to summarize institutional characteristics and safety compliance. Comparisons of staffing patterns by time period were evaluated using inferential tests, with statistical significance set at p < 0.05. RESULTS: The response rate was 100% (47/47). The mean number of MRI units per hospital was 5.32 ± 3.50, and 3.0 T systems were predominant. Compliance was universal for closed-circuit television (CCTV) monitoring and MR Conditional monitoring devices but was low for ferromagnetic detection systems (17.0%), Magnetic Resonance Safety Officer (MRSO) appointment (17.0%), and Level 1 safety education (23.4%). MRI-related safety incidents were reported by 44.7% of hospitals. Two-person staffing decreased significantly from 97.9% on weekday daytime hours to 55.3% on Saturdays, 44.7% at night, and 38.3% on holidays. The overall composite safety maturity score was 63.08 ± 15.04. DATA CONCLUSION: Considerable variability remains in MRI safety management across Korean tertiary hospitals. These findings highlight the need for national MRI safety standards in the Republic of Korea. EVIDENCE LEVEL: 4. TECHNICAL EFFICACY: Stage 4.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Measuring and evaluating safety competence in MRI environments among healthcare professionals: A systematic review.Radiography (London, England : 1995) · 2026
- MRI-related safety incidents during acute cerebrovascular care: A descriptive analysis of a national incident-reporting database in Japan.Clinical neurology and neurosurgery · 2026
- Enhancing Inpatient MRI Safety and Workflow Using Standardized MRI Care Level Guidelines and Electronic Medical Record-Integrated Dashboard.Academic radiology · 2026
- MRI Safety and Disaster Management for MR-Simulator and MR-Linacs.Seminars in radiation oncology · 2026
- Patient Satisfaction with MRI Department Service and Staff in Public and Private Hospitals in Aseer, Saudi Arabia: A Cross-Sectional Study.Patient preference and adherence · 2026
- Data-driven Assessment of Magnetic Resonance Imaging (MRI) Utilization and Aggregate Electrical Demand in U.S. Healthcare Facilities.Academic radiology · 2026