Compressed Sensing-Accelerated Cartesian Real-Time Phase Contrast MRI With Shared Velocity Encoding: A Clinical Performance Evaluation.
Gang Yin, Yan Liang, Mi Li, Xinling Yang, Lin Chen, Kai Yang, Chen Cui, Xiuyu Chen,
Imaging Center, Fuwai Hospital and National Center for Cardiovascular Diseases, Peking Union Medical College and Chinese Academy of Medical Sciences, Beijing, China. · Medical Imaging Center, Shandong Second Provincial General Hospital, Jinan, Shandong, China.
Key points
From the paper's conclusion- RS acquisition time was 16.0 ± 2.3 s; total imaging time including breath-hold instructions was 26.0 ± 2.3 s, whereas CSRT acquisition time was 10.0 s, yielding a statistically significant difference in scan duration. All hemodynamic parameters demonstrated strong significant correlation between CSRT PC and RS PC MRI (r ≥ 0.89) and good-to-excellent inter-method agreement (ICC ≥ 0.87). CSRT PC yielded significantly lower values for peak velocity (bias = -16.72 ± 11.48 cm/s [-13.31% ± 9.68%]), peak flow rate (bias = -36.28 ± 35.06 mL/s [-9.18% ± 8.98%]), and peak mean velocity (bias = -6.49 ± 7.06 cm/s [-11.02% ± 13.14%]) relative to the RS, but net flow rate was not significantly different between methods (bias = 0.15% ± 0.60 L/min [4.16% ± 17.28%, p = 0.058]).
Abstract
BACKGROUND: Conventional segmented phase contrast (PC) MRI is the clinical reference standard for cardiovascular hemodynamic quantification, but prolonged acquisition times, mandatory breath-holding, and susceptibility to arrhythmia limit its utility in vulnerable populations. PURPOSE: To validate the clinical performance of compressed sensing-accelerated real-time (CSRT) PC MRI with shared velocity encoding (SVE) for fast, free-breathing cardiac flow assessment. STUDY TYPE: Prospective. POPULATION: 63 patients (46 male/17 female, mean age 51.5 ± 15.1 years) with diverse cardiac disease. FIELD STRENGTH/SEQUENCE: 3.0 T; single-shot RF-spoiled gradient echo sequence with SVE (free-breathing CSRT PC MRI) and velocity encoded segmented gradient echo sequence (breath-hold routine segmented (RS) PC MRI). ASSESSMENT: Hemodynamic parameters-peak velocity, peak flow rate, net flow rate, and peak mean velocity-were measured at the sinotubular junction plane for both CSRT and RS datasets. Scan time of the two sequences was recorded. STATISTICAL TESTS: Pearson/Spearman correlation analysis, intraclass correlation coefficient (ICC), Bland-Altman analysis, paired t-test/Wilcoxon test. A p value < 0.05 was considered significant. RESULTS: RS acquisition time was 16.0 ± 2.3 s; total imaging time including breath-hold instructions was 26.0 ± 2.3 s, whereas CSRT acquisition time was 10.0 s, yielding a statistically significant difference in scan duration. All hemodynamic parameters demonstrated strong significant correlation between CSRT PC and RS PC MRI (r ≥ 0.89) and good-to-excellent inter-method agreement (ICC ≥ 0.87). CSRT PC yielded significantly lower values for peak velocity (bias = -16.72 ± 11.48 cm/s [-13.31% ± 9.68%]), peak flow rate (bias = -36.28 ± 35.06 mL/s [-9.18% ± 8.98%]), and peak mean velocity (bias = -6.49 ± 7.06 cm/s [-11.02% ± 13.14%]) relative to the RS, but net flow rate was not significantly different between methods (bias = 0.15% ± 0.60 L/min [4.16% ± 17.28%, p = 0.058]). DATA CONCLUSIONS: CSRT PC provided satisfactory agreement with conventional RS PC while shortening scan time and enabling free-breathing acquisition. EVIDENCE LEVEL: 1. TECHNICAL EFFICACY: Stage 1.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of Compressed Sensing Accelerated Magnetization Prepared Rapid Acquisition Gradient Echo for Fast and Reliable Brain Volume Measurement.Tomography (Ann Arbor, Mich.) · 2026
- Volumetric real-time MRI for the guidance of cardiac catheterization at 0.55T.Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance · 2026
- Whole-Brain Cerebral Microvascular Pulsatility Imaging Using Dynamic pCASL MRI With Segmented Stack-of-Stars Golden-Angle Radial Acquisition.Magnetic resonance in medicine · 2026
- Motion-Robust Approach for 4D Cardiac Cine Using 2D Real-Time Acquisitions and Slice-to-Volume Reconstruction.Magnetic resonance in medicine · 2026
- Patch-Based Diffusion Reconstruction for Accelerated Cardiac CinearXiv · 2026
- PROSPECTIVE COMPARISON OF OPTIMIZED COMPRESSED SENSING AND GRAPPA-ACCELERATED 3D-TOF-MRA FOR EVALUATION OF UNTREATED INTRACRANIAL ANEURYSMS AT 1.5 T: A NON-INFERIORITY STUDYmedRxiv · 2026