Ex Vivo Comparative Experiments Demonstrating the Ablation Effect of a Temperature-Controlled Radiofrequency Ablation System.
Kenjiro Yamamoto, Takao Itoi, Yuki Joyama, Takayoshi Tsuchiya, Ryosuke Tonozuka, Shuntaro Mukai, Kazumasa Nagai, Hiroyuki Kojima,
Department of Gastroenterology and Hepatology, Tokyo Medical University, Tokyo, Japan. · Department of Pathology, Akita University Hospital, Akita, Japan.
Key points
From the paper's conclusion- Therefore, this ex vivo study aimed to directly compare the ablation characteristics of the TC‐RFA with those of a power‐controlled RFA (PC‐RFA).
Abstract
BACKGROUND: A temperature-controlled radiofrequency ablation (TC-RFA) system has been introduced to reduce excessive thermal damage and enable safer treatment. However, the ablation effect of this system remains to be fully elucidated. Therefore, this ex vivo study aimed to directly compare the ablation characteristics of the TC-RFA with those of a power-controlled RFA (PC-RFA). METHODS: Pig liver specimens were used. The TC-RFA using all four probe types and the PC-RFA were performed step-by-step to macroscopically evaluate the ablation area and speed. Linear regression models with interaction terms were used to compare changes in ablation dimensions over time between RFA devices and probes. RESULTS: The TC-RFA produced different short-axis ablation diameters depending on the probe even at the same temperature setting and generated an ablation effect at an earlier time, with substantially less change in ablation dimensions over time than the PC-RFA. In contrast, the PC-RFA had the slowest onset, but its ablation area continued to increase in width and depth with longer ablation durations. These differences in ablation behavior over time were statistically significant (Holm-adjusted p < 0.001). CONCLUSIONS: In these exploratory ex vivo experiments, the TC-RFA system produced earlier and more consistent ablation effects over time than the PC-RFA system.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Efficacy and safety of a self-expandable stent-based endobiliary radiofrequency ablation: A preclinical study.Bioengineering & translational medicine · 2026
- Assessment of temperature dynamics in pulsed field ablation with a variable-loop circular catheter: a comparative analysis of waveform configurations and irrigation rates in specimens of bovine ventricular myocardiumEP Europace · 2025
- Limited benefit of half-normal vs normal saline irrigation for augmenting current delivery with open-irrigated temperature-controlled ablation.Heart rhythm O2 · 2026
- Assumption-bounded reduced electro-thermal modeling of lesion formation in radiofrequency cardiac ablation: power-time allocation, thermal latency, perfusion, and temperature limitation.Computer methods in biomechanics and biomedical engineering · 2026
- From Ionic Agitation to Precision Neuroablation: A Bioengineering Perspective on the Mechanisms, Modalities, and Clinical Applications of Radiofrequency Technology in Pain Management.Bioengineering (Basel, Switzerland) · 2026
- Thermal Ablation for Autonomously Functioning Thyroid Nodules: A Prospective Comparison of Radiofrequency and Microwave Ablation.Endocrine practice : official journal of the American College of Endocrinology and the American Association of Clinical Endocrinologists · 2026