A modified transosseous quadriceps tendon refixation technique demonstrates advantageous primary stability: A biomechanical investigation.
Nora S Michel, Lucas K Palma Kries, Andre Frank, Michael J Raschke, J Christoph Katthagen, Simon Oeckenpöhler, Christoph Kittl, Adrian Deichsel,
Department of Trauma, Hand and Reconstructive Surgery University Hospital Münster Münster Germany. · Faculty of Engineering and Computer Science Osnabrück University of Applied Sciences Osnabrück Germany.
Key points
From the paper's conclusion- The modified transosseous technique with anterior drill holes yields higher failure loads and less elongation than the conventional TL method, with biomechanical performance comparable to SA fixation, supporting its use as a viable alternative.
Abstract
PURPOSE: Recent biomechanical research suggests suture anchor (SA) refixation may be superior to conventional transosseous longitudinal (TL) repair for quadriceps tendon ruptures. This study aimed to evaluate whether a modified transosseous technique with anteriorly angled drill holes enhances primary stability compared to the aforementioned techniques. METHODS: Twenty-one human cadaveric patellae were randomly assigned to three groups (n = 7): SA (two SAs), TL (three longitudinal transosseous drill holes) and TA (three anteriorly angled 45° transosseous drill holes). Identical suture material (Ethibond #2) and a four-strand suture configuration were used in all groups. Prior to fixation, computed tomography (CT) scans measured bone mineral density (BMD) and cortical thickness. Biomechanical testing simulated early postoperative loading using a universal testing machine: preload (20 N for 30 s), cyclic loading (500 cycles, 20-100 N at 1 Hz) and a load-to-failure test (20 mm/s). As primary outcome, ultimate load to failure was recorded; secondarily, cyclic elongation and stiffness were measured. One-way analysis of variance (ANOVA) with Bonferroni-corrected post hoc t tests was used for statistical analysis. RESULTS: The TA group had a significantly higher ultimate load to failure than the TL group (472.7 ± 22.4 vs. 344.8 ± 108.8 N, p < 0.05). No significant differences were found between the SA group (438.9 ± 77.4 N) and the other groups. The TL group showed significantly greater cyclic elongation (6.22 ± 1.03 mm) compared to the SA group (4.42 ± 0.82 mm, p < 0.01). The TA group (5.03 ± 0.73 mm) showed no significant difference compared to either group. Stiffness did not differ significantly (p > 0.05). CONCLUSION: The modified transosseous technique with anterior drill holes yields higher failure loads and less elongation than the conventional TL method, with biomechanical performance comparable to SA fixation, supporting its use as a viable alternative. LEVEL OF EVIDENCE: N/A.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Biomechanical comparison of innovative suture techniques for repair of Achilles tendon rupture in a porcine flexor tendon model.Archives of orthopaedic and trauma surgery · 2026
- Comparison of an anchor plate with suture anchor and transosseous constructs for rotator cuff repair in a polyurethane foam block: a biomechanical studyBMC Musculoskeletal Disorders · 2026
- Bioplating demonstrates increasing micromovement during cyclic loading after anterior closed wedge high tibial osteotomy despite no significant difference in failure load compared to angle-stable plate fixation.Journal of experimental orthopaedics · 2026
- An In Vitro Biomechanical Study of Quadriceps Tendon Reconstruction with Synthetic Mesh: A Novel TechniqueBiomechanics · 2026
- Open Proximal Rectus Femoris Tendon Full-Thickness Tear Repair With Double-Row Suture Anchor Transosseous-Equivalent Construct With Biologic Augmentation Using Bone Marrow Aspirate Concentrate and Bovine Bioinductive Collagen Implant.Arthroscopy techniques · 2026
- Greater deformation and lower stiffness, but similar ultimate failure strength with fixed-loop cortical suspensory device compared to metal interference screw for femoral fixation of bone-patellar tendon-bone grafts in anterior cruciate ligament reconstruction.Journal of experimental orthopaedics · 2026