Novel Target for Pathogenic Leptospira Detection via Colorimetric Loop-Mediated Isothermal Amplification (LAMP): Development, Validation, and Comparative Analysis.
Peyman Ghoraishizadeh, Benjie Clemente, Kween Saimuang, Sakaorat Lertjuthaporn, Sukhonta Limsampan, Patcharapan Suwannin, Mongkol Techakasikornpanich, Galayanee Doungchawee,
Center for Research Innovation and Biomedical Informatics, Faculty of Medical Technology, Mahidol University, Nakhon Pathom, Thailand. · Department of Medical Technology, Faculty of Pharmacy, University of Santo Tomas, Manila, Philippines.
Key points
From the paper's conclusion- The LAMP-orpH assay offers a rapid, sensitive, and specific tool for early leptospirosis diagnosis. Its simple colorimetric readout and minimal equipment requirements make it ideal for use in rural and resource-limited settings, providing a practical alternative to MAT, conventional PCR, and qPCR.
Abstract
BACKGROUND: Leptospirosis is a globally prevalent zoonotic disease that remains challenging to diagnose due to its nonspecific symptoms and the limitations of current diagnostic methods and targets. To address these challenges, we developed and validated a naked-eye loop-mediated isothermal amplification (LAMP) assay targeting the newly identified orpH region for the rapid detection of pathogenic Leptospira species. METHODS: Comparative genomic analysis identified orpH as a conserved sequence among pathogenic Leptospira. A colorimetric LAMP assay (LAMP-orpH) was developed and evaluated using reference strains and serum samples obtained from patients with suspected leptospirosis. The diagnostic performance of the assay on these serum samples was evaluated in comparison with conventional PCR and the microscopic agglutination test (MAT), with the real-time PCR assay targeting lipL32 designated as the reference standard. RESULTS: The LAMP-orpH assay demonstrated 87.2% sensitivity and 94.8% specificity among 76 suspected cases and 29 controls, showing strong agreement with real-time PCR (κ = 0.826). It outperformed MAT (63.8% sensitivity) and conventional PCR (23.4% sensitivity), achieving a detection limit of 102 cells/mL in spiked serum. CONCLUSION: The LAMP-orpH assay offers a rapid, sensitive, and specific tool for early leptospirosis diagnosis. Its simple colorimetric readout and minimal equipment requirements make it ideal for use in rural and resource-limited settings, providing a practical alternative to MAT, conventional PCR, and qPCR.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Hybrid Molecular Lateral Flow Assay Targeting the lipL32 Gene for Rapid and Highly Sensitive Detection of Pathogenic Leptospira.Tropical medicine and infectious disease · 2026
- Evaluation of Novel Primers for Rapid Single-Step Detection of Pathogenic Leptospira spp. Using Duplex Polymerase Chain Reaction (PCR) Assay.Tropical life sciences research · 2026
- Clinical evaluation of an in-house probe-based real-time PCR assay for early diagnosis of leptospirosis compared with Microscopic Agglutination TestGalle Medical Journal · 2026
- Optimization and clinical validation of new and improved TaqMan Real-Time PCR assays for the detection of pathogenic Leptospira.medRxiv · 2026
- Taxonomic identification and characterization of Leptospira from retrospective clinical samples in the PhilippinesmedRxiv · 2026
- Evaluation of the Leptocheck-WB IgM rapid test for acute leptospirosis among acute undifferentiated fever patients in western Uganda: very low sensitivity against locally circulating Leptospira serogroupsmedRxiv · 2026