Spatially Controlled Dual-Functionalization of a Titanium Keratoprosthesis: A Proof-of-Concept In Vivo Study.
Silvia González Gómez, Carla Caballero, Justin Christopher D'Antin, Anton Barraquer, Gemma Julio, Francisco Javier Gil, José María Manero, Rafael I Barraquer
Biomaterials, Biomechanics and Tissue Engineering Group, Department of Materials Science and Engineering, Universitat Politècnica de Catalunya. Barcelona Tech (UPC), Barcelona East School of Engineering (EEBE), Barcelona, Spain. · Centro de Oftalmología Barraquer, Barcelona, Spain.
Key points
From the paper's conclusion- These findings support the technical feasibility of spatially controlled dual-functionalization and provide preliminary descriptive observations of short-term tissue-implant interactions during the first postoperative month, when complications are most frequent.
Abstract
Keratoprosthesis (KPro), an artificial corneal implant, is an established surgical option for restoring vision in severe corneal blindness; however, long-term outcomes remain limited by complications such as fibrosis and infection. Current biomaterial strategies generally lack the ability to address these distinct pathological processes at spatially defined implant-tissue interfaces. Here, we present a modified titanium KPro backplate with spatially controlled dual-functionalization, enabling spatially decoupled functionalization of opposing implant surfaces to address different biological challenges. The study was designed as an exploratory proof-of-concept and was not intended to determine the independent contribution or efficacy of either surface functionalization. The anterior surface was coated with antimicrobial silver nanoparticles, while the posterior surface was coated with a P17-derived peptide construct designed for surface immobilization and intended to target fibrosis-associated pathways. Side-specific surface modification was supported by scanning electron microscopy, fluorescence microscopy, and x-ray photoelectron spectroscopy. A proof-of-concept in vivo study was performed in a rabbit model over 1 month, comparing dual-functionalized and uncoated modified KPro devices. Corneal melting occurred in 3/6 control and 3/5 treated eyes; among eyes with melting, the mean number of affected corneal sectors was 3 and 1, respectively. Mean graft opacification was 60% in controls and 40% in treated eyes, and mean iris adhesion was 85% in control eyes and 70% in treated. Scanning electron microscopy qualitatively showed variable amounts of ECM-like fibrillar material and cellular-like structures at the implant interfaces, with no consistent qualitative difference between groups. Clinical findings compatible with infection were observed in 2/6 control eyes and 0/5 treated eyes. These findings support the technical feasibility of spatially controlled dual-functionalization and provide preliminary descriptive observations of short-term tissue-implant interactions during the first postoperative month, when complications are most frequent. This approach provides a basis for further evaluation of spatially programmed KPro surface engineering.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- NIR-Activatable Implant Nanocoating: Spatiotemporally Controlled Drug Release, Photothermal Bactericidal Activity, and Peritoneal Adhesion Suppression.ACS applied materials & interfaces · 2026
- Amyloid-like Protein-Metal Sulfide Nanocoatings for Synergistic Photothermal and Antibacterial Implant Surface ProtectionACS Nano · 2025
- A sustained living coating for infectious keratitis therapy via probiotic consumption and production effectsScience Advances · 2026
- Development of an Injectable Antibacterial Adhesive Sealant for Sutureless Repair of Corneal Injuries.ACS applied bio materials · 2026
- Case Report: A novel surgical drainage bag-derived plastic sheet-assisted keratoprosthesis implantation technique: a retrospective case seriesFrontiers in Bioengineering and Biotechnology · 2026