Probe-Based Near-Infrared Autofluorescence for Parathyroid Identification During Thyroid Surgery: Clinical Outcomes, Surgeon Confidence, and Autofluorescence Thresholds.

Anthony Y Cheung, Bo Wang, Yoshiyuki Saito, Marika D Russell, Justin M Soffer, Okenwa C Okose, Lily N Trinh, Adithya Srikanthan,

Division of Thyroid and Parathyroid Endocrine Surgery, Department of Otolaryngology-Head and Neck Surgery, Massachusetts Eye and Ear, Harvard Medical School, Boston, Massachusetts, USA. · Department of Otolaryngology-Head and Neck Surgery, Massachusetts Eye and Ear, Harvard Medical School, Boston, Massachusetts, USA.

Observational

Key points

From the paper's conclusion
  • PTeye provides adjunctive confirmation of suspected parathyroid tissue, including glands initially identified with low or moderate visual confidence.

Abstract

BACKGROUND: We evaluated experience with probe-based near-infrared autofluorescence (NIRAF) system (PTeye) in parathyroid gland (PTG) identification during total thyroidectomy. METHODS: In this retrospective study, patients undergoing total thyroidectomy using PTeye (2020-2022; n = 84) were compared with a historical visual-identification cohort (2018-2019; n = 103). RESULTS: PTG identification rates were comparable, although 42% of PTGs detected with PTeye were initially visualized with low or moderate confidence. Significantly fewer patients in the PTeye cohort had PTGs incidentally identified on histology (4.8% vs. 16.5%, p = 0.018), but there was no difference in the incidence of autotransplanted PTGs or total inadvertent parathyroidectomy. Postoperative calcium was higher with PTeye (median 9.2 vs. 9.0 mg/dL, p < 0.001). Postoperative PTH, transient, and permanent hypoparathyroidism did not differ. An autofluorescence ratio cutoff of 2.5 improved specificity compared with the manufacturer-recommended cutoff of 1.2. CONCLUSIONS: PTeye provides adjunctive confirmation of suspected parathyroid tissue, including glands initially identified with low or moderate visual confidence.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Probe-Based Near-Infrared Autofluorescence for Parathyroid Identification During Thyroid Surgery: Clinical Outcomes, Surgeon Confidence, and Autofluorescence Thresholds — Science Skim