TrustSpec : Simulation‐Grounded Physics‐Driven Spectral Inversion With Uncertainty Quantification for SiC Epitaxial Thickness Metrology
Zijie Nie, Shunlin Yang, Qingjie Guo
Key points
From the paper's conclusion- In a reproducible simulation study over 0.4–25 μm, TrustSpec achieves a mean absolute error (MAE) of 0.110 μm and remains effective below 1 μm, where a classical FFT pipeline fails.
Abstract
Accurate thickness metrology is essential for silicon carbide (SiC) epitaxial layers, yet classical Fourier‐transform infrared (FTIR) fringe analysis relies on artifact‐specific preprocessing, degrades below the fringe‐resolution limit, and provides no calibrated confidence for individual measurements. This paper presents TrustSpec, a simulation‐grounded, physics‐driven spectral inversion framework intended as a proof of concept for trustworthy SiC epitaxial thickness metrology. A dispersive two‐interface interference model with randomized instrument disturbances is used as a spectral digital twin to generate labeled training spectra, and the same twin is reused at inference time as an analysis‐by‐synthesis referee. A deep ensemble estimates thickness from a physics‐conjugate Fourier‐magnitude optical‐thickness representation; the ensemble sample standard deviation and a windowed re‐synthesis consistency ratio define a two‐channel trust plane for release, review, or rejection. In a reproducible simulation study over 0.4–25 μm, TrustSpec achieves a mean absolute error (MAE) of 0.110 μm and remains effective below 1 μm, where a classical FFT pipeline fails. Split‐conformal recalibration yields 94.6% empirical coverage for a nominal 95% interval. The combined trust policy releases 90.5% of in‐distribution spectra at an MAE of 0.066 μm and rejects all 600 tested simulated 28–38 μm thickness‐extrapolation spectra; a 25–28 μm test shows a gradual rejection transition near the training boundary. Channel ablations, referee‐window sensitivity, shifted‐nuisance stress tests, and model‐mismatch experiments delimit the scope of these simulation results and motivate batch residual monitoring, reference‐wafer checks, and periodic two‐angle acquisition for drift detection. Validation on measured SiC wafers remains necessary before production deployment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Neural-Assisted Triangulation Framework for Broadband Optical Monitoring of Multilayer CoatingsCoatings · 2026
- Denoising and Baseline Correction of Low-Scan FTIR Spectra: A Benchmark of Deep Learning Models Against Traditional Signal ProcessingBioengineering · 2026
- Uncertainty Aware AFM-IR Hyperspectral Cube Reconstruction via Convolutional-Attentive Conditional Neural ProcessesarXiv · 2026
- Physics-Based Drude-Lorentz Synthetic Spectra and Transformer-Based Reconstruction of Broadband Absorption Coefficients from Sparse Wavelength MeasurementsJournal of Physics: Conference Series · 2026
- Deep Learning-Based Prediction of Tool Influence Function for Nanometric Control in Space Optical MaterialJournal of Astronomy and Space Sciences · 2026
- Uncertainty-aware Bayesian neural networks for FTIR gas-phase process monitoring: a semi-synthetic feasibility studyat - Automatisierungstechnik · 2026