Multi‑Channel CE‑SDS: Core‑Performance Evaluation, System Suitability Test, and Method Validation in Biopharmaceutical Analysis.
Xinying Zhao, Dongke Liu, Xikai Wen, Xiaoxia Zhang, Hongmei Tang, Tingting Li, Hongxu Chen
School of Bioengineering, Beijing Polytechnic University, Beijing, China. · SCIEX, Beijing, China.
Key points
From the paper's conclusion- As their application in biopharmaceutical characterization expands, standardized system suitability test (SST) and method validation protocols tailored to the multi‐channel CE platform are urgently needed.
Abstract
Capillary electrophoresis-sodium dodecyl sulfate (CE-SDS) is the gold standard for the purity assessment of protein-based biopharmaceuticals, yet traditional single-capillary CE-SDS workflows suffer from inherent throughput limitations. Commercial multi-capillary CE platforms deliver higher analytical throughput and represent powerful tools for high-throughput biopharmaceutical testing. As their application in biopharmaceutical characterization expands, standardized system suitability test (SST) and method validation protocols tailored to the multi-channel CE platform are urgently needed. Herein, we evaluated the core performance of a multi-channel CE platform and reported the first scientific and efficient CE-SDS SST and method validation scheme for multi-channel CE platform. First, the core performance of an 8-channel CE platform was systematically evaluated, including analytical consistency with the well-validated single-channel reference CE system, intra-instrument consistency across eight capillaries, and the absence of inter-capillary crosstalk. SST and comprehensive CE-SDS method validation were implemented concurrently, and all SST parameters met predefined acceptance criteria throughout the method validation sequence. The CE-SDS method validation covered specificity, linear range, limit of quantification (LOQ), accuracy, and precision, all performed in compliance with ICH requirements. The whole workflow was completed in only a few days, fully leveraging the high-throughput advantage of multi-channel CE platform while ensuring robust and reliable analytical results. The protocol and experimental data herein deliver practical utility and valuable regulatory-aligned guidance for peer laboratories and biopharmaceutical enterprises to deploy multi-channel CE in early-stage biologic screening, process development, quality release, and other scenarios.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Recent advances and applications of capillary electrophoresis-sodium dodecyl sulfate for characterization and fragment identification of monoclonal antibodies and their derivativesThe Analyst · 2025
- Validation of a High-Throughput Fluorescent Capillary Electrophoresis Sodium Dodecyl Sulfate Method for Monoclonal Antibody Size Heterogeneity Assessment.Electrophoresis · 2026
- Sensitive and Robust Capillary Electrophoretic Analysis of Host Cell Residual DNA in Biopharmaceutical Samples.Analytical chemistry · 2026
- Key strategic considerations for pooling multiple drug substance batches in biopharmaceutical drug product manufacturing.Journal of pharmaceutical sciences · 2026
- Continuous sampling/transfer strategy to overcome sampling-volume mismatch in heart-cutting two-dimensional liquid chromatography: Application to ertapenem impurity profiling.Journal of chromatography. A · 2026
- Intact Proteoform Analysis by Capillary Electrophoresis–Mass Spectrometry. Are We There Yet?Angewandte Chemie International Edition · 2025