Unexpected Peaks in Reduced Capillary Gel Electrophoresis Associated With Hair and Dust Contamination.
Jeremy D Osko, Shannon Rivera, Mark A Haverick, Lauren B Favell, Thomas P Niedringhaus, Joop C M Waterval
Large Molecule Analytical, Merck & Co., Inc., Rahway, New Jersey, USA. · Pharmaceutical Analysis & Digital Technologies, Merck & Co., Inc., Rahway, New Jersey, USA.
Key points
From the paper's conclusion- Here, we explored the impact of hair and dust contamination in CGE across three different monoclonal antibodies (mAbs) designated as mAb 1, mAb 2, and mAb 3.
Abstract
Capillary gel electrophoresis (CGE) is a size-based analytical technique used for monitoring the critical quality attribute of molecular fragmentation. The ability to distinguish protein degradation from method artifacts remains an industry-wide challenge, driving the need to better understand contributing factors of atypical peak formations. One relatively uncharacterized contributor is contamination introduced during sample preparation. Despite standard operating protocols, analytical method protocols, and implementation of best practices, it is still possible to unintentionally introduce contamination. Two known yet subtle sources of relevant contaminants are hair from the analyst performing the experiment and dust present within the laboratory. Here, we explored the impact of hair and dust contamination in CGE across three different monoclonal antibodies (mAbs) designated as mAb 1, mAb 2, and mAb 3. Through a series of practical experimentation assays, including contaminant spiking experiments and timecourse assessments, six new peaks were identified with a direct correlation to the introduction of hair and dust contaminants. The six peaks were highly reproducible under reducing conditions but not observed in identical tests under nonreducing conditions. This suggests a reducing agent, such as beta-mercaptoethanol, is required for the formation of Peaks 1-6. Proteolytic activity was also assessed through enzymatic digestion and activity assays. Our data found no evidence of protease involvement, suggesting the peaks are not a result of protein degradation. The relative abundances of corresponding light chains and heavy chains remained consistent, whereas new abundances were observed for Peaks 1-6. One potential suspect was keratin, a fibrous protein that serves as the foundation for hair and could be present in dust. Peptide mapping confirmed an increase in abundance of keratin peptides for samples treated with hair compared to untreated controls. This work highlights a realistic application where the unintentional introduction of hair or dust into CGE samples could produce additional peaks.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Sensitive and Robust Capillary Electrophoretic Analysis of Host Cell Residual DNA in Biopharmaceutical Samples.Analytical chemistry · 2026
- Evaluation of Pre-Analytical Performance Using IFCC Quality Indicators, Six Sigma Metrics and Root Cause Analysis in a Biochemistry Laboratory-A Retrospective Study.Diagnostics (Basel, Switzerland) · 2026
- From Enrichment to Insight: Leveraging Preparative cIEF Enrichments and Characterization to Support Data‐Driven Integration Practices for cIEF‐based AnalyticsELECTROPHORESIS · 2026
- Capillary gel electrophoresis analysis of the forced degradation study products of a therapeutic monoclonal antibody.Journal of pharmaceutical and biomedical analysis · 2026
- Continuous sampling/transfer strategy to overcome sampling-volume mismatch in heart-cutting two-dimensional liquid chromatography: Application to ertapenem impurity profiling.Journal of chromatography. A · 2026
- Validation of a High-Throughput Fluorescent Capillary Electrophoresis Sodium Dodecyl Sulfate Method for Monoclonal Antibody Size Heterogeneity Assessment.Electrophoresis · 2026