A Novel Left‐Hand Tool Control System for Endoscopic Submucosal Dissection: Technical Description and Initial Clinical Experience (With Video)
Hidetaka Ohta, Atsushi Miwa
Key points
From the paper's conclusion- We developed a left‐hand tool control system, assembled from a polytetrafluoroethylene tube, fixation tape, two plastic one‐touch clips serving as a movable device grip, and a sterile adapter cap, that relocates the accessory‐channel inlet into the left‐hand control field so that the knife is advanced and withdrawn with the left hand while the right hand maintains continuous scope control.
Abstract
In conventional endoscopic submucosal dissection (ESD), the operator's right hand alternates between stabilizing the endoscope shaft and manipulating the treatment device. We developed a left‐hand tool control system, assembled from a polytetrafluoroethylene tube, fixation tape, two plastic one‐touch clips serving as a movable device grip, and a sterile adapter cap, that relocates the accessory‐channel inlet into the left‐hand control field so that the knife is advanced and withdrawn with the left hand while the right hand maintains continuous scope control. Unlike existing left‐hand techniques, the grasping point lies only a short distance from the relocated inlet, which prevents the sheath from buckling during advancement. We report the initial experience in 41 consecutive ESD procedures (three esophageal, 13 gastric, and 25 colorectal) performed between June 2025 and April 2026. In every procedure, the operator maintained shaft torque and moved the knife simultaneously and without assistance. En bloc and R0 resection were each achieved in 39 procedures (95.1%), with no muscle injury or intraprocedural perforation. Within 3 weeks, one colorectal procedure was followed by delayed bleeding treated endoscopically and another by an unscheduled 3‐day prolongation of hospital stay. In an exploratory comparison adjusted for specimen area, fibrosis, age, and organ, ESD time was not shorter than in conventional procedures by the same operator during the preceding 2 years (time ratio 1.23, 95% confidence interval 1.00–1.52). The system was feasible in this single‐operator series; its effect on procedural efficiency and its safety require prospective evaluation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development of a Coaxial Over‐the‐Scope Bimanual Robot Enabling Lesion‐Aligned Manipulation for ESDInternational Journal of Medical Robotics and Computer Assisted Surgery · 2026
- Helping Hand: A Continuum Robotic Grasper for Traction in Single-Channel Endoscopic Submucosal DissectionarXiv · 2026
- Colorectal ESD from basics to latest innovations!Best practice & research. Clinical gastroenterology · 2026
- Endoscopic Hand Suturing After Gastric Endoscopic Submucosal Dissection in Patients with Low to Intermediate Risk of Delayed Bleeding: A Retrospective Pilot Study.Digestive diseases and sciences · 2026
- Undersaline ESD for Large Colorectal Lesions: A Narrative Review.Medicina (Kaunas, Lithuania) · 2026
- Efficacy of Endoscopic Hand Suturing for Prevention of Adverse Events After Gastric Endoscopic Submucosal Dissection: A Multicenter Propensity Score-Matched Study.Endoscopy · 2026