A Novel Left‐Hand Tool Control System for Endoscopic Submucosal Dissection: Technical Description and Initial Clinical Experience (With Video)

Hidetaka Ohta, Atsushi Miwa

Key points

From the paper's conclusion
  • We developed a left‐hand tool control system, assembled from a polytetrafluoroethylene tube, fixation tape, two plastic one‐touch clips serving as a movable device grip, and a sterile adapter cap, that relocates the accessory‐channel inlet into the left‐hand control field so that the knife is advanced and withdrawn with the left hand while the right hand maintains continuous scope control.

Abstract

In conventional endoscopic submucosal dissection (ESD), the operator's right hand alternates between stabilizing the endoscope shaft and manipulating the treatment device. We developed a left‐hand tool control system, assembled from a polytetrafluoroethylene tube, fixation tape, two plastic one‐touch clips serving as a movable device grip, and a sterile adapter cap, that relocates the accessory‐channel inlet into the left‐hand control field so that the knife is advanced and withdrawn with the left hand while the right hand maintains continuous scope control. Unlike existing left‐hand techniques, the grasping point lies only a short distance from the relocated inlet, which prevents the sheath from buckling during advancement. We report the initial experience in 41 consecutive ESD procedures (three esophageal, 13 gastric, and 25 colorectal) performed between June 2025 and April 2026. In every procedure, the operator maintained shaft torque and moved the knife simultaneously and without assistance. En bloc and R0 resection were each achieved in 39 procedures (95.1%), with no muscle injury or intraprocedural perforation. Within 3 weeks, one colorectal procedure was followed by delayed bleeding treated endoscopically and another by an unscheduled 3‐day prolongation of hospital stay. In an exploratory comparison adjusted for specimen area, fibrosis, age, and organ, ESD time was not shorter than in conventional procedures by the same operator during the preceding 2 years (time ratio 1.23, 95% confidence interval 1.00–1.52). The system was feasible in this single‐operator series; its effect on procedural efficiency and its safety require prospective evaluation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A Novel Left‐Hand Tool Control System for Endoscopic Submucosal Dissection: Technical Description and Initial Clinical Experience (With Video) — Science Skim