Inter-software variability of Centiloid quantification in amyloid PET: A head-to-head comparison of five software platforms.
Jun Zhang, Brian J Burkett, Christopher G Schwarz, Nur Dizdar, Matthew P Johnson, Cem Bilgin, Bradley J Kemp, Jolanta M Durski,
Department of Radiology Mayo Clinic Rochester Minnesota USA. · Department of Quantitative Health Sciences Mayo Clinic Rochester Minnesota USA.
Key points
From the paper's conclusion- Agreement was excellent (intraclass correlation coefficient [ICC] 0.969; 95%CI 0.961-0.975). Some platforms produced systematically higher CL values versus others slightly lower. LoA reached ± 30 CL. Diagnostic accuracy was high (area under the curve [AUC]: 0.949-0.975), with sensitivity 0.923-0.968 and specificity 0.667-0.806.
Abstract
INTRODUCTION: Centiloid (CL) scaling standardizes amyloid positron emission tomography (PET) quantification across tracers and platforms; however, variability across software implementations may affect diagnostic classification. This study evaluated inter-software variability and diagnostic performance across five platforms using identical 1 8F-florbetapir datasets. METHODS: Retrospectively, 192 patients undergoing 1 8F-florbetapir PET/computed tomography (CT) and magnetic resonance imaging (MRI) were analyzed. CL values were generated using four US Food and Drug Administration (FDA) -cleared platforms and an in-house Centiloid standard pipeline. Agreement was assessed using intraclass correlation coefficient, with bias and limits of agreement evaluated by linear modeling and Bland-Altman analysis. Diagnostic performance was assessed using receiver operating characteristic (ROC) analysis and classification against visual interpretation. RESULTS: Agreement was excellent (intraclass correlation coefficient [ICC] 0.969; 95%CI 0.961-0.975). Some platforms produced systematically higher CL values versus others slightly lower. LoA reached ± 30 CL. Diagnostic accuracy was high (area under the curve [AUC]: 0.949-0.975), with sensitivity 0.923-0.968 and specificity 0.667-0.806. DISCUSSION: Despite excellent agreement, systematic differences persist and may affect classification near thresholds, supporting consistent use of a single processing pipeline.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Cross-platform Comparison of Quantitative Florbetapir Amyloid PET Using MIMneuro and Neurophet SCALE PET.Clinical nuclear medicine · 2026
- Centiloid Calibration and Application of MUSE-Based Amyloid PET Quantification for Florbetapir (AV-45)medRxiv · 2026
- From image reconstruction to clinical prognosis: A comprehensive analysis of centiloid-based amyloid PET assessmentEuropean Journal of Nuclear Medicine and Molecular Imaging · 2026
- Quantitative Amyloid Brain Imaging: A Literature Review of the Centiloid Scale in Alzheimer Disease EvaluationJournal of Nuclear Medicine Technology · 2026
- Amyloid-PET pipeline choice influences classification of preclinical Alzheimer's diseasemedRxiv · 2026
- Concordance between amyloid PET and CSF biomarkers in clinical setting: a cross-platform comparison and in-depth analysis of discordant casesScientific Reports · 2025