Elecsys pT181p immunoassay: Clinical performance for ruling out amyloid pathology.
Craig Ritchie, Sayuri Hortsch, Clara Quijano-Rubio, Patrick Menzel, Imke Kirste, David Caley, Anna D Burke, Christopher C Rowe,
Scottish Brain Sciences, 3rd floor Scottish Gas Murrayfield Stadium South Stand Edinburgh UK. · Roche Diagnostics GmbH Penzberg Bavaria Germany.
Key points
From the paper's conclusion- At a care setting-agnostic cutoff of 0.934 pg/ml, high negative predictive value (NPV) (93.8%) and acceptable positive predictive value (PPV) (46.6%) were observed against amyloid-positron emission tomography (PET). A primary care-specific cutoff of 0.722 pg/ml showed high NPV (97.9%) against amyloid-PET. Amyloid positivity prevalence for each cutoff population was 22.5% and 13.1%, respectively.
Abstract
INTRODUCTION: Plasma biomarker rule-out tests can be used early in the Alzheimer's disease (AD) diagnostic pathway to identify people with low likelihood of amyloid pathology. METHODS: A care setting-agnostic cutoff for the Elecsys® Phospho-Tau (181P) plasma (pT181p) immunoassay was determined and validated in two independent, diverse cohorts, representative of real-world clinical practice (N = 604; N = 787, respectively). An additional cutoff was assessed in a subset of people reflective of primary care (n = 312). RESULTS: At a care setting-agnostic cutoff of 0.934 pg/ml, high negative predictive value (NPV) (93.8%) and acceptable positive predictive value (PPV) (46.6%) were observed against amyloid-positron emission tomography (PET). A primary care-specific cutoff of 0.722 pg/ml showed high NPV (97.9%) against amyloid-PET. Amyloid positivity prevalence for each cutoff population was 22.5% and 13.1%, respectively. DISCUSSION: The pT181p immunoassay is a high-performance test that may substantially help improve access to amyloid pathology testing for people requiring confirmation, potentially leading to timely clinical decisions and better outcomes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluating plasma phosphorylated tau217 for clinical rule-out of cerebral amyloid pathology: Evidence from a memory clinic cohort.Journal of Alzheimer's disease : JAD · 2026
- Elecsys pTau217 plasma immunoassay detection of amyloid pathology in clinical cohortsAlzheimer s & Dementia · 2026
- Clinical performance and decision thresholds of an automated plasma p-tau217 assay for amyloid pathology in a Taiwanese memory clinic.Journal of neurology · 2026
- Clinical utility of LUMIPULSE plasma p-tau217 to identify amyloid positivity among candidates for amyloid-targeting treatments in a real-world cohort.Alzheimer's & dementia (Amsterdam, Netherlands) · 2026
- Head-to-head comparison of plasma biomarker assays across platforms for amyloid pathology in a multicenter cohort.Science bulletin · 2026
- A prospective observational study measuring the stability of plasma pTau181 using the Elecsys Phospho-Tau (181P) plasma immunoassay for early detection of Alzheimer's disease.Alzheimer's research & therapy · 2026