Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions

Superior Vena Cava Compromise During Transcatheter Atrial Septal Defect Closure: A Lesson in Implantation Geometry.

Pernthaler Maximilian, Nagel Marina, Nikolaus Haas, André Jakob

Abteilung für Kinderkardiologie und Pädiatrische Intensivmedizin, LMU Klinikum, LMU Medizin, Ludwig-Maximilians-Universität München, Munich, Germany.

Case report

Key points

From the paper's conclusion
  • We report the case of a 5-year-old girl with a multifenestrated aneurysmal secundum ASD.

Abstract

Transcatheter closure procedure of multifenestrated aneurysmal atrial septal defects (ASDs) remains technically challenging: Optimal device selection and positioning should achieve complete closure of all fenestrations while accommodating the complex three-dimensional atrial septal anatomy and its relationship to adjacent cardiac structures. Because superior vena cava (SVC) obstruction is exceedingly rare, it may easily be overlooked unless caval venous inflow is systematically evaluated by transesophageal echocardiography (TEE) after device deployment following a pre-release TEE checklist. We report the case of a 5-year-old girl with a multifenestrated aneurysmal secundum ASD. A Uni-occluder was not available; therefore, a patent foramen ovale (PFO) occluder was used for defect closure, which resulted in near-complete obstruction of the SVC by the right atrial disc of the device. The complication was recognized after positioning by intraprocedural TEE, prompting device retrieval prior to release. Subsequent closure with a dedicated ASD occluder achieved complete closure of all fenestrations without compromising caval venous return. This case highlights the importance of a concept we refer to as implantation geometry: the interplay between device-specific design, the implantation site, and complex three-dimensional septal anatomy. Recognition and careful assessment of this relationship will help optimize procedural planning and improve procedural safety in selected patients with complex, multifenestrated ASDs. Comprehensive three-dimensional anatomical assessment, appropriate device selection, and systematic TEE evaluation of the cavoatrial junction before device release are essential for safe transcatheter closure of multifenestrated ASDs-even when defect sizing appears appropriate. In particular, the geometric characteristics of PFO occluders, including their relatively large right atrial disc, should be considered when assessing the risk of SVC obstruction.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Superior Vena Cava Compromise During Transcatheter Atrial Septal Defect Closure: A Lesson in Implantation Geometry — Science Skim