Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions

Zero-Contrast Transcatheter Aortic Valve-in-Valve Replacement Guided by Continuous Hemodynamic Monitoring.

Mizuki Miura, Yusuke Watanabe, Jo Omiya, Ken Kozuma

Division of Cardiology, Teikyo University School of Medicine, Tokyo, Japan.

Key points

From the paper's conclusion
  • Continuous pressure monitoring may support intraprocedural assessment and guide valve optimization when contrast exposure is being avoided in patients with renal dysfunction.

Abstract

An 89-year-old man with structural deterioration of a surgical aortic bioprosthesis and renal dysfunction underwent zero-contrast transcatheter aortic valve-in-surgical valve (TAV-in-SAV) implantation. A SavvyWire provided continuous left ventricular pressure monitoring and rapid pacing during the procedure. After implantation of a 23-mm self-expanding valve, a residual transvalvular gradient was detected, prompting balloon postdilatation under rapid pacing. Repeat invasive hemodynamic assessment showed improvement, and echocardiography demonstrated a mean gradient of 12 mmHg. At 12 months, the patient was in New York Heart Association class I, with preserved renal function and a mean gradient of 8 mmHg. Continuous pressure monitoring may support intraprocedural assessment and guide valve optimization when contrast exposure is being avoided in patients with renal dysfunction.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Zero-Contrast Transcatheter Aortic Valve-in-Valve Replacement Guided by Continuous Hemodynamic Monitoring — Science Skim