Advanced science (Weinheim, Baden-Wurttemberg, Germany)

Suppressing Photothermal Artifacts in Flexible Fluorescent Fiber-Optic Sensors for Accurate and Precise In Situ Temperature Monitoring.

Sijie Su, Jinjian Yu, Pengyuan Dong, Xin Wang, Jing Lyu, Jiulin Gan, Enhai Song, Qinyuan Zhang

State Key Laboratory of Luminescent Materials and Devices, Guangdong Provincial Key Laboratory of Fiber Laser Materials and Applied Techniques, South China University of Technology, Guangzhou, China. · Guangdong Energy Group Science and Technology Research Institute Co., Ltd., Guangzhou, China.

Key points

From the paper's conclusion
  • Here, we propose a partial-inner-cladding optical fiber (PICOF) to suppress photothermal artifacts by isolating the sensing layer from the coupling interface and the fiber core, where self-heating predominantly occurs.

Abstract

Ratiometric flexible fluorescent fiber-optic sensors, featuring self-calibration and precise spatial localization, are emerging as a robust platform for in situ thermal monitoring, advancing applications such as battery safety management and biomedical diagnostics. However, measurement artifacts induced by the inherent photothermal effect remain largely overlooked, posing a challenge to high-accuracy temperature sensing. Here, we propose a partial-inner-cladding optical fiber (PICOF) to suppress photothermal artifacts by isolating the sensing layer from the coupling interface and the fiber core, where self-heating predominantly occurs. Compared with conventional fiber structures, the PICOF extends the allowable excitation power from ≤ 15 to 61 mW and improves the luminescence signal-to-noise ratio from 36.6 to 42 dB, enabling high-fidelity readout with a relative standard deviation below 0.09% even under power fluctuations and deformation disturbances. Consequently, the LuVO4:25%Yb3+/Er3+-composite PICOF achieves record-high accuracy (maximum absolute error δmax = 0.104°C), precision (standard deviation σ = 0.05°C), and temperature resolution (TR = 0.2°C) among reported fluorescent fiber-optic sensors. Furthermore, practical demonstrations underscore its utility for early warning of thermal hazards and incipient failure in sodium-ion batteries, and as an in situ diagnostic reference for the Bárány caloric test. This work provides an effective and generalizable strategy for the advancement of high-accuracy fluorescent thermometry.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Suppressing Photothermal Artifacts in Flexible Fluorescent Fiber-Optic Sensors for Accurate and Precise In Situ Temperature Monitoring — Science Skim