A Wearable Microfluidic Sweat-Patch Pilot Study for Detecting Breast-Cancer Proteins CEA and PIP/GCDFP-15.
Fatemeh Esmaeildoost, Soroosh Arshadi, Fatemeh Rahimi, Mehrdad Najafpour, Milad Falahat Chian, Eleftherios Sachlos, Siu Ning Leung, Razieh Salahandish
Laboratory of Advanced Biotechnologies for Health Assessments (Lab-HA), Lassonde School of Engineering, York University, Toronto, Ontario, Canada. · Department of Biology, Faculty of Science, York University, Toronto, Ontario, Canada.
Key points
From the paper's conclusion- These findings represent, to our knowledge, the first paired sweat-blood comparison of CEA and PIP using a passive capillary-driven wearable patch and establish analytical requirements, including sample volume, analyte recovery, assay sensitivity, and biomarker selection, for noninvasive sweat-based breast cancer monitoring.
Abstract
Blood-based measurement of carcinoembryonic antigen (CEA) and gross cystic disease fluid protein-15 (GCDFP-15; prolactin-inducible protein, PIP) remains the standard for breast cancer monitoring but is invasive. We present WearDOXXN+, a capillary-driven microfluidic patch for noninvasive sweat collection to investigate paired sweat-blood analysis. Ten participants wear the patch during exercise; matched sweat and blood samples are analyzed by enzyme-linked immunosorbent assay (ELISA). The platform collects sweat within 30 min, with 100% success. CEA was detected in sweat and showed a disease-stage-associated trend; however, concentrations in healthy participants and survivors remained below the assay's lower limit of quantification, highlighting analytical sensitivity as a limitation. PIP was measurable in blood, with the highest level observed in Stage III disease, although several values exceeded the assay's validated upper limit, while remaining undetectable in sweat. Validation showed the patch recovers 65% of loaded fluid, and recovered eluate retains loaded CEA concentration (recovery 108%, 92%, and 98%), confirming sample handling does not deplete the analyte. These findings represent, to our knowledge, the first paired sweat-blood comparison of CEA and PIP using a passive capillary-driven wearable patch and establish analytical requirements, including sample volume, analyte recovery, assay sensitivity, and biomarker selection, for noninvasive sweat-based breast cancer monitoring.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Wearable Lateral Flow Patch for Noninvasive Sweat Protein MonitoringACS Sensors · 2026
- Agricultural waste-derived sweat sensor integrated into wettability-tunable microfluidics for noninvasive metabolic evaluation.The Analyst · 2026
- Wearable Sweat Biosensors for Inflammatory Biomarkers (CRP, IL-6 and TNF-α): A Systematic Review of Analytical and Clinical Evidence.Biosensors · 2026
- Wearable Microfluidic Chip Integrated with Chrono-Sampling for Cortisol Detection in Sweat.Analytical chemistry · 2026
- Deep learning-based noninvasive assessment of additively manufactured biomechanical models for sports training simulationNondestructive Testing And Evaluation · 2026
- Continuous sweat biomarker monitoring for early detection of sacral pressure injuries in postoperative patients: A prospective dual-zone wearable sensor study.Journal of tissue viability · 2026