Cardiac substructure dose in thoracic VMAT without explicit constraints: FMO versus SPO on uRT-TPOIS with cross-platform reference to eclipse PO.

Chun Wu, Zhenmei Cao, Tingting Duan, Lixiang Han, Bin Wang, Weijie Lei

Department of Radiation Oncology, Xiamen Cancer Hospital, Xiamen, China. · Department of Radiation Oncology, Second Affiliated Hospital of Xuzhou Medical University, Xuzhou, China.

Key points

From the paper's conclusion
  • Whether explicit cardiac substructure constraints provide incremental benefit remains to be tested in dedicated within‐platform studies.

Abstract

BACKGROUND: Cardiac substructure doses are increasingly recognized as more specific predictors of radiation-induced cardiac toxicity than whole-heart metrics, yet these substructures are often left unconstrained in routine thoracic volumetric-modulated arc therapy (VMAT) planning due to contouring burden or low prioritization. PURPOSE: This study aims to evaluate, in the absence of explicit cardiac substructure constraints, whether optimizer selection is associated with differences in cardiac substructure doses within a single treatment planning system (TPS), and to provide a descriptive cross-system reference against a widely used clinical platform. METHODS: Thirty patients (10 each across esophageal, breast, and non-small cell lung cancers) initially treated with clinically delivered Photon Optimizer (PO) plans on the Eclipse platform were retrospectively replanned using two optimization algorithms on the uRT-TPOIS platform: Fluence Map Optimization (FMO) and Stochastic Platform Optimization (SPO). No explicit dose constraints were applied to any cardiac substructure. Prespecified primary endpoints were mean heart dose (MHD), mean dose to the left anterior descending artery (LAD Dmean), mean dose to the left ventricle (LV Dmean), and mean lung dose (MLD), with the latter serving as a pulmonary toxicity reference. All other dosimetric parameters were analyzed as secondary/exploratory endpoints. Paired comparisons used the Wilcoxon signed-rank test, with Benjamini-Hochberg false discovery rate correction applied separately within each cancer type and pairwise comparison for primary endpoints. Secondary endpoints were exploratory and not adjusted for multiplicity. RESULTS: Intra-platform comparison within uRT-TPOIS showed that SPO was associated with lower MHD, LAD Dmean, and MLD than FMO in esophageal cancer, and lower MHD, LAD Dmean, and LV Dmean in breast cancer (q < 0.05). In NSCLC, no significant differences were observed for the primary endpoints. In secondary analyses, SPO was associated with shorter beam-on time across all sites and improved target coverage in NSCLC. In descriptive cross-system comparisons with Eclipse PO, SPO was associated with lower primary endpoint doses in esophageal and breast cancers, whereas in NSCLC, only MHD differed significantly, favoring PO. Exploratory analyses suggested additional differences in other cardiac substructures, but these were not adjusted for multiplicity and should be considered hypothesis-generating. CONCLUSION: Under the specific uRT-TPOIS protocol and without explicit cardiac substructure constraints, SPO was associated with lower primary endpoint doses than FMO in esophageal and breast cancers, but not in NSCLC. Cross-system comparisons were descriptive and confounded by platform-level factors. Whether explicit cardiac substructure constraints provide incremental benefit remains to be tested in dedicated within-platform studies.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Cardiac substructure dose in thoracic VMAT without explicit constraints: FMO versus SPO on uRT-TPOIS with cross-platform reference to eclipse PO — Science Skim