Impact of tissue heterogeneity on intracranial Cesium-131 brachytherapy dosimetry: A patient-specific Monte Carlo Study.
Tirthraj Adhikari, Jae W Jung, Sherif M Gadoue, Puja Sharma, Tomas Montegegro, Farhad Jafari, Clara Ferreira
University of Minnesota, Minneapolis, Minnesota, USA. · East Carolina University, Greenville, North Carolina, USA.
Key points
From the paper's conclusion- Although the TG-43 formalism provides accurate dose estimates in homogeneous media, it introduces clinically meaningful local and global dosimetric deviations in heterogeneous intracranial environments. Careful consideration is required when reporting doses to the target and OARs, particularly given the potential for substantially higher doses to the irradiated skull in certain cases (up to 3.2-fold in this study), as well as the possibility of target underdosage within the high-risk clinical target volume (HR-CTV). Patient-specific MC modeling highlights the importance of incorporating tissue heterogeneity into intracranial brachytherapy dose calculations to improve dosimetric accuracy and optimize treatment planning.
Abstract
BACKGROUND: AAPM TG-43-based dose calculations for intracranial cesium-131 (131Cs) brachytherapy ignore patient-specific heterogeneities, which can affect dose accuracy. This study investigates the dosimetric impact of these heterogeneity effects using Monte Carlo (MC) simulations and patient-specific imaging data. PURPOSE: To quantify the impact of tissue heterogeneity on dose distributions in intracranial brachytherapy using 131Cs sources and to evaluate the accuracy of AAPM TG-43-based treatment planning system (TPS) calculations. METHODS: TG-43 dosimetric parameters for the 131Cs source were independently characterized and benchmarked against published consensus data. MC simulations using MCNP were performed in both homogeneous water phantoms and heterogeneous voxelized lattice phantoms derived from CT DICOM RT data. Single and multiple-seed configurations were investigated. TPS and MC dose distributions were compared using dose-volume histogram (DVH) metrics such as (V60, V100, D2cc, and mean dose), isodose distributions, dose difference maps, and local and global dose difference analyzes. Spatial agreement was quantified using gamma analysis. RESULTS: Strong agreement was observed between TPS and MC calculations in homogeneous media. However, heterogeneous patient-specific MC simulations revealed systematic differences in DVH metrics for the target volume and organs at risk, including the skull and skin. Local dose differences revealed pronounced deviations near high-density bony regions, where dose enhancement effects were observed, while global dose difference analyzes demonstrated an overall dosimetric bias. Gamma pass rates decreased in heterogeneous regions, consistent with these spatial dose discrepancies. Compared with the MC simulations, the TG-43-based TPS overestimated the mean dose to the target by approximately 23.63 ± 4.97 % , V60 and V100 by approximately 24.42 ± 4.97 % , and 23.75 ± 4.97 % , respectively, and the mean dose to the skin by approximately 73.60 ± 4.97 % , while underestimating the mean dose to the skull by 220.90 ± 4.97 % . CONCLUSION: Although the TG-43 formalism provides accurate dose estimates in homogeneous media, it introduces clinically meaningful local and global dosimetric deviations in heterogeneous intracranial environments. Careful consideration is required when reporting doses to the target and OARs, particularly given the potential for substantially higher doses to the irradiated skull in certain cases (up to 3.2-fold in this study), as well as the possibility of target underdosage within the high-risk clinical target volume (HR-CTV). Patient-specific MC modeling highlights the importance of incorporating tissue heterogeneity into intracranial brachytherapy dose calculations to improve dosimetric accuracy and optimize treatment planning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Radiobiological dosimetry using hybrid imaging-derived spatiotemporal dose distributions for131I therapy of metastatic thyroid cancer.Physics in medicine and biology · 2026
- Investigation of TG-43 dose calculation algorithm for volumetric HDR Brachytherapy using Monte Carlo simulation in TOPAS.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- A Monte Carlo based dose verification tool for an HDR Ir-192 brachytherapy source and its evaluation on extremity malignancies.Journal of applied clinical medical physics · 2026
- Dosimetric comparison of HDR Co-60 and Ir-192 brachytherapy sources by model-based dose calculation algorithms and its clinical implication in different brachytherapy sites and techniques.Brachytherapy · 2026
- High-fidelity Monte Carlo dosimetry of radiosensitive organs in head and neck CT: A virtual TLD study with comprehensive statistical uncertainty analysis.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Clinical test cases for commissioning, QA, and benchmarking of model‐based dose calculation algorithms in¹ ⁹² Ir HDR gynecologic tandem and ring brachytherapyMedical Physics · 2026